Fig 29 - Position of the tonometer A900 (SL9900)
4.9
HOW TO MEASURE THE INTRAOCULAR PRESSURE
Tell the patient to keep the eyes open during the exam.
The measurement time should be as short as possible.
1
Ask the patient to open and close the eyes for a few seconds,
so that the tear film is evenly distributed on the corneal
surface.
2
Choose the blue filter to activate the fluorescein.
3
Move forward the slit lamp until the measuring prism is into
contact with the center of the cornea.
4
Place the graduated scale on the value 1.
5
Observing the applanated surface with the microscope you will
see two semicircles of the same size. The two semicircles can
have different sizes depending on the ocular pressure. When
the tonometer is in the right position, the two semicircles have
the same dimension.