DEVICE USE
4.1
HOW TO INSTALL THE TONOMETER Z800 ON THE SLIT
LAMP
1
Install the tonometer support on top of the microscope, on its
compartment. Block it with the screw on the device.
2
Place the tonometer on the support pivot.
Fig 13 - How to place the tonometer support
Fig 14 - How to place the tonometer Z800
3
Free the arm fro the protection support and move it from the
resting position to the measurement position.
4
If necessary adjust the measuring prism support position to
align it with the microscope shooting channel. Block the
tonometer by means of the screws placed under the
measuring scale knob.