![CSO A900 Instructions For Use Manual Download Page 48](http://html.mh-extra.com/html/cso/a900/a900_instructions-for-use-manual_2692899048.webp)
INSTRUCTIONS FOR USE
Z8A9F9 | IFU323EN00
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
46/60
Fig 36 - Drying the measuring prism
Fig 37 - Measuring prisms container
5.3.1
MEASURING PRISM DISINFECTION
While disinfecting the measuring prism use disposable gloves.
1
Fill a clean and disinfected container with the disinfectant
solution. The disinfectant solution shall cover the measuring
prisms completely.
2
Place the measuring prisms in the container.
3
The measuring prisms shall not be on top of each other.
4
The measuring prisms shall be completely immersed in the
disinfectant solution
5
Let the disinfectant have its effect following the instructions of
the disinfectant solution Producer.