5.4
TONOMETER FUNCTIONING CHECK
CAUTION
If the device is outside from the indicated calibration tolerances,
contact the Technical Service for reparation and calibration.
To check the device correct functioning you need to use the specific
functioning check accessory.
1
Place the functioning check accessory on the device.
2
The functioning check accessory has 5 markings on its body.
The markings correspond to the values indicated on the
measuring knob: 0 (central), 2, 6 (ends). The functioning check
accessory longer end shall be placed facing the examiner.
Fig 42 - How to place the functioning check
accessory on the tonometer Z800
Fig 43 - How to place the functioning check
accessory on the tonometer A900 and F900