Fluorescein semicircle too wide or too small.
Image
Cause
Solution
The image of the two
semicircles is too wide. The
intraocular pressure
measurement might result
higher than the real one.
Carefully clean the measuring
prism before positioning it
again.
The image of the two
semicircles is too small. The
intraocular pressure
measurement might result
lower than the real one.
Carefully clean the measuring
prism before positioning it
again.
The measuring prism does not touch the cornea or an excessive
pressure has been applied.
Image
Cause
Solution
If the patient draws his/her
head back slightly, irregular
movements of the eye will
happen. this will cause an
unstable contact between
measuring prism and eye.
Ask the patient to stay perfectly
still and open the eyes better.
Repeat the measurement.