Technical Discussion
106
Automatic Calibration
Because light absorption by hemoglobin is wavelength dependent and
because the mean wavelength of LEDs varies, an oximeter must know the
mean wavelength of the sensor's red LED to accurately measure SpO
2
. During
manufacturing, the mean wavelength of the red LED is encoded in the sensor.
During monitoring, the N-560's software uses the coefficients that are
appropriate for the wavelength of that individual sensor's red LED; these
coefficients are then used to determine SpO
2
. The N-560’s software obtains
coefficients when the N-560 is turned on, periodically thereafter, and each
time a new sensor is connected.
Additionally, to compensate for differences in tissue thickness, the light
intensity of the sensor's LEDs is adjusted automatically.
Functional Versus Fractional Saturation
This N-560 measures functional saturation -- oxygenated hemoglobin
expressed as a percentage of the hemoglobin that can transport oxygen. It
does not detect significant amounts of dysfunctional hemoglobin, such as
carboxyhemoglobin or methemoglobin. In contrast, hemoximeters such as the
IL482 report fractional saturation -- oxygenated hemoglobin expressed as a
percentage of all measured hemoglobin, including measured dysfunctional
hemoglobins. To compare functional saturation measurements to those from
an instrument that measures fractional saturation, fractional measurements
must be converted as follows:
Measured versus Calculated Saturation
When saturation is calculated from a blood gas partial pressure of oxygen
(PO
2
), the calculated value may differ from the SpO
2
measurement of a pulse
oximeter. This usually occurs because the calculated saturation was not
appropriately corrected for the effects of variables that shift the relationship
between PO
2
and pH, temperature, the partial pressure of carbon dioxide
(PCO
2
), 2,3-DPG, and fetal hemoglobin. See Figure 46.
functional
saturation
=
fractional
saturation
100
-
(
%
carboxyhemoglobin
+
%
methemoglobin
)
x
100
Summary of Contents for Nellcor OxiMax N-560
Page 1: ...Service Manual OxiMaxTM Technology Nellcor OxiMax N 560 Pulse Oximeter TM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 128: ...Technical Discussion 118...
Page 130: ...Technical Discussion 120...
Page 132: ...Technical Discussion 122...
Page 134: ...Technical Discussion 124...
Page 136: ...Technical Discussion 126...
Page 138: ...Technical Discussion 128...
Page 139: ...Technical Discussion N 560 Service Manual 129 Figure 55 Main PCB Parts Locator Diagram...
Page 140: ...Technical Discussion 130...
Page 142: ...Technical Discussion 132...
Page 144: ...Technical Discussion 134...
Page 146: ...Technical Discussion 136...
Page 147: ...Technical Discussion N 560 Service Manual 137 Figure 59 Front Panel Parts Locator Diagram...
Page 148: ...Technical Discussion 138...
Page 149: ...Technical Discussion N 560 Service Manual 139 Figure 60 Power Supply Schematic Diagram...
Page 150: ...Technical Discussion 140...
Page 151: ...Technical Discussion N 560 Service Manual 141 Figure 61 Power Supply Parts Locator Diagram...
Page 152: ...Technical Discussion 142...
Page 156: ......
Page 157: ......