Technical Discussion
N-560 Service Manual
109
O
XI
M
AX
Technology
The N-560 is designed to use Nellcor brand sensors containing
O
XI
M
AX
technology. These sensors can be identified by their deep blue plug color. All
O
XI
M
AX
sensors contain a memory chip carrying information about the sensor
which the oximeter needs for correct operation, including the sensor’s
calibration data, model type, troubleshooting codes, and error detection data.
This unique oximetry architecture enables several new features with the
N-560.
When an
O
XI
M
AX
sensor is connected to the N-560, the N-560 will first read
the information in the sensor memory chip, check it to make sure that there
are no errors, and then load the data to begin monitoring. As the N-560 reads
the information, it flashes the sensor model number on its display. This
process only takes a couple of seconds. Once the reading process is complete,
the sensor model number will stop flashing on the display, and then the N-560
will begin monitoring. The sensor model number disappears once the N-560
starts tracking the patient’s SpO
2
and pulse rate.
Monitors containing
O
XI
M
AX
technology, including the N-560, use calibration
data contained in the sensor in calculating the patient’s SpO
2
. By having the
calibration in the sensor, rather than the N-560, Nellcor is able to improve the
published accuracy of many sensors, because the calibration coefficients can
be tailored to each sensor. Consult the accuracy card included with the N-560
for specific accuracy information for the N-560 with different Nellcor
sensors.
The N-560 uses the information in the
O
XI
M
AX
sensor to tailor troubleshooting
messages for the clinician. The sensor contains coding that tells the N-560
what kind of sensor is being used. When deciding what messages to display,
the N-560 takes into account the sensor type and recommended patient site for
that model sensor. The N-560
O
XI
M
AX
system therefore has an intelligent
troubleshooting system.
NELL-3 Interface
The NELL-3 interfaces with the N-560 through two connections. Figure 47
illustrates the interconnection, and Table 21 and Table 22 identify the signals.
Summary of Contents for Nellcor OxiMax N-560
Page 1: ...Service Manual OxiMaxTM Technology Nellcor OxiMax N 560 Pulse Oximeter TM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 128: ...Technical Discussion 118...
Page 130: ...Technical Discussion 120...
Page 132: ...Technical Discussion 122...
Page 134: ...Technical Discussion 124...
Page 136: ...Technical Discussion 126...
Page 138: ...Technical Discussion 128...
Page 139: ...Technical Discussion N 560 Service Manual 129 Figure 55 Main PCB Parts Locator Diagram...
Page 140: ...Technical Discussion 130...
Page 142: ...Technical Discussion 132...
Page 144: ...Technical Discussion 134...
Page 146: ...Technical Discussion 136...
Page 147: ...Technical Discussion N 560 Service Manual 137 Figure 59 Front Panel Parts Locator Diagram...
Page 148: ...Technical Discussion 138...
Page 149: ...Technical Discussion N 560 Service Manual 139 Figure 60 Power Supply Schematic Diagram...
Page 150: ...Technical Discussion 140...
Page 151: ...Technical Discussion N 560 Service Manual 141 Figure 61 Power Supply Parts Locator Diagram...
Page 152: ...Technical Discussion 142...
Page 156: ......
Page 157: ......