Audible Alarm Settings and Service Functions
N-560 Service Manual
35
When the desired option is indicated in the %SpO
2
display, press the
SatSeconds
Alarm Limit button to initiate the current selection.
Note
: Menu items greater than 5 cannot be accessed when a valid sensor is
connected to the N-560.
Service Menu Options
Service menu options can be accessed only when the sensor is disconnected
from the N-560. Only qualified service personnel should access these options.
Refer to
on page 33 for instructions on how to access
the menu options and make selections within them.
Menu Item 6 (In-Sensor Trend Mode)
This menu item is used to set up the type of event data that a single use
O
XI
M
AX
sensor will record within the sensor. Three options can be accessed
by first pressing the
SatSeconds
Alarm Limit button when menu item 6 is
displayed. This function is only applicable to
O
XI
M
AX
sensors that do not have
any event data recorded in the sensor. If the
O
XI
M
AX
sensor has event data
recorded in it, the N-560 will record event data in the same format as the
recorded data.
Use the
Adjust Up
or
Adjust Down
button to select the desired option.
• Option 0 = record SpO
2
event data
• Option 1 = record SpO
2
plus pulse rate event data
• Option 2 = do not use in-sensor record function
When the desired option is indicated in the %SpO
2
display, press the
SatSeconds
Alarm Limit button to initiate the current selection.
Menu Item 7 (RS-232 Nurse Call Polarity)
This menu item is used to setup the polarity of the RS-232 Nurse Call signal
available at the N-560 rear panel data port. Two options can be accessed by
first pressing the
SatSeconds
Alarm Limit button when menu item 7 is
displayed.
Use the
Adjust Up
or
Adjust Down
button to select the desired option.
Summary of Contents for Nellcor OxiMax N-560
Page 1: ...Service Manual OxiMaxTM Technology Nellcor OxiMax N 560 Pulse Oximeter TM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 128: ...Technical Discussion 118...
Page 130: ...Technical Discussion 120...
Page 132: ...Technical Discussion 122...
Page 134: ...Technical Discussion 124...
Page 136: ...Technical Discussion 126...
Page 138: ...Technical Discussion 128...
Page 139: ...Technical Discussion N 560 Service Manual 129 Figure 55 Main PCB Parts Locator Diagram...
Page 140: ...Technical Discussion 130...
Page 142: ...Technical Discussion 132...
Page 144: ...Technical Discussion 134...
Page 146: ...Technical Discussion 136...
Page 147: ...Technical Discussion N 560 Service Manual 137 Figure 59 Front Panel Parts Locator Diagram...
Page 148: ...Technical Discussion 138...
Page 149: ...Technical Discussion N 560 Service Manual 139 Figure 60 Power Supply Schematic Diagram...
Page 150: ...Technical Discussion 140...
Page 151: ...Technical Discussion N 560 Service Manual 141 Figure 61 Power Supply Parts Locator Diagram...
Page 152: ...Technical Discussion 142...
Page 156: ......
Page 157: ......