Responding to System Alarms
Tr
oubleshooting
Force FX Electrosurgical Generator C User’s Guide
7-11
69–71
80
Software malfunction.
Turn off, then turn on the generator. If the
alarm number reappears, record the number
and call the Covidien Service Center.
81
Internal component
malfunction.
Do not attempt to use the generator. Record
the number and call the Covidien Service
Center.
90
95
Microcontroller malfunction.
Contact your Biomedical Engineering
Department.
100–105
110–119
Software malfunction.
Turn off, then turn on the generator. If the
alarm number reappears, record the number
and call the Covidien Service Center.
120
Calibration malfunction.
Contact your Biomedical Engineering
Department.
121
Software malfunction.
Turn off, then turn on the generator. If the
alarm number reappears, record the number
and call the Covidien Service Center.
122
123–126
Calibration malfunction.
Microcontroller malfunction.
Contact your Biomedical Engineering
Department.
130–134
136–138
150
Software malfunction.
Turn off, then turn on the generator. If the
alarm number reappears, record the number
and call the Covidien Service Center.
151
Microcontroller malfunction.
Contact your Biomedical Engineering
Department.
152
Software malfunction.
Turn off, then turn on the generator. If the
alarm number reappears, record the number
and call the Covidien Service Center.
154
Microcontroller malfunction.
Contact your Biomedical Engineering
Department.
160
Internal component
malfunction.
Do not attempt to use the generator. Record
the number and call the Covidien Service
Center.
Number
Description
Recommended Action