Responding to System Alarms
Tr
oubleshooting
Force FX Electrosurgical Generator C User’s Guide
7-13
190
Bipolar up arrow, bipolar
down arrow,
and/or a bipolar mode button
(Precise, Standard, Macro) may
be stuck.
1. Turn off, then turn on the generator. Do not
press buttons or accessory activation
devices during the self-test.
2. If the alarm number reappears, disconnect
all accessories. Then, turn off and turn on
the generator again.
If the alarm number reappears, record the
number and call the Covidien Service
Center.
191
Cut up arrow, cut down
arrow, and/or a cut mode
button (Low, Pure, Blend) may
be stuck.
192
Coag up arrow, coag down
arrow, and/or a coag mode
button (Desiccate, Fulgurate,
Spray) may be stuck.
193
Recall button may be stuck.
194
Handswitch or Monopolar 1
Footswitch cut pedal may be
stuck.
195
Handswitch or Monopolar 1
Footswitch coag pedal may be
stuck.
196
Handswitch or Monopolar 2
Footswitch cut pedal may be
stuck.
197
Handswitch or Monopolar 2
Footswitch coag pedal may be
stuck.
198
Handswitch or Bipolar
Footswitch pedal may be
stuck.
199–203
Internal diagnostics or
microcontroller malfunction.
Contact your Biomedical Engineering
Department.
206–207
Software malfunction.
Turn off, then turn on the generator. If the
alarm number reappears, record the number
and call the Covidien Service Center.
208–209
Microcontroller malfunction.
Contact your Biomedical Engineering
Department.
Number
Description
Recommended Action