Monopolar Coag Controls
2-6
Force FX Electrosurgical Generator C User’s Guide
Monopolar Coag Controls
Coag indicator
When you activate coag, this
indicator illuminates blue and
an activation tone sounds.
Power buttons
Press
Δ
to increase
the power.
Press
∇
to decrease
the power.
Mode indicators
Illuminates green
when you press the
corresponding mode
button.
Spray mode button
Select to evenly coagulate a wide
area of tissue with a spray of sparks;
penetration is less deep and tissue
area is larger than in fulgurate mode.
Fulgurate mode button
Select to fulgurate an area of tissue with a spray of
sparks. This unit is equipped with an additional
fulgurate mode which incorporates a lower crest factor
(LCF) than the factory default fulgurate mode.
Details about this additional fulgurate mode, LCF
fulgurate, are provided in Chapter 4,
This is the default monopolar coag mode. However, the
default coag mode can be changed to either Desiccate
or Spray through the serial port on the rear panel;
contact your institution’s Biomedical Engineering
Department.
Desiccate mode button
Select to desiccate the area of
tissue that is in direct contact
with the active electrode.
Coag display
Shows the power setting,
in watts, for the selected
mode.