Chapter 9 - Lung diffusing capacity - 87
DLCO without apnoea (“intrabreath” method)
1. Remove the sampling line from the flowmeter
2. Select
Single Breath CO diffusing capacity
from the
Test
menu
3. Confirm or in case change the test options:
- Breath hold time (deselect the check box)
- Simulated test (it allows to carry out the manoeuvre without inhaling gas, useful
for teaching purposes)
- Hb (haemoglobin)
- Anatomic dead space
4. Wait for the zeroing of the analyser is completed and plug in again the sampling
line in the flowmeter
5. Connect the Patient to the mouthpiece and let her/him breathing at rest until the
message “Carry out Vital Capacity” is prompted
6. Carry out a Slow Vital Capacity if desired (this is not necessary for completing the
DLCO test)
7. Invite the Patient to exhale up to Residual Volume and press
F2
during the
maximal expiration
8. Coach the Patient to inhale up to Total Lung Capacity and to expire slowly
maintaining the flow rates within the two target lines shown on the real-time
window until the Vital Capacity has been exhaled.
A dashed line indicates the 90% of the Best Vital Capacity (if a FVC or SVC test is
available in the archive, otherwise it indicates the 90% of the Predicted VC), in order to
verify the correctness of the manoeuvre.
Double-click with the left mouse button in order to modify the test options (wash-out
volume, sampling volume, calculation of the real apnoea time…).
Multiple breath DLCO test (“Steady-State” method)
In this test, the patient breathes regularly during the whole test (about 8 min): graphs
and results will be displayed breath by breath.
Normally, during the first 2 minutes the patient inhales ambient air, then such a mixture:
0.1% CO, 0.1% CH4, 21% O2, N2 balance.
1. Remove the sampling line from the flowmeter
2. Select
CO diffusing capacity steady state
from the
Test
menu
4. Wait for the zeroing of the analyser is completed and plug in again the sampling
line in the flowmeter
5. Connect the Patient to the mouthpiece and let her/him breathing at rest until the
message “Carry out Vital Capacity” is prompted
6. Carry out a Slow Vital Capacity if desired (this is not necessary for completing the
DLCO test)
7. While the patient is exhaling, press
F2
in order to start the mixture
8. When the steady state is reached, press
F3
to end the test.
Right-clicking you can generate new graphs and parameters to be displayed, both during
the test and at the end.
Summary of Contents for Quark PFT
Page 11: ...Getting started ...
Page 27: ...Measurements ...
Page 32: ...32 Quark PFT User Manual ...
Page 33: ...Installation ...
Page 42: ...42 Quark PFT User Manual ...
Page 43: ...Calibration ...
Page 50: ...50 Quark PFT User Manual ...
Page 51: ...Database Management ...
Page 57: ...Spirometry ...
Page 72: ...72 Quark PFT User Manual ...
Page 73: ...Lung Volumes ...
Page 78: ...78 Quark PFT User Manual ...
Page 79: ...P0 1 Respiratory drive ...
Page 82: ...82 Quark PFT User Manual ...
Page 83: ...Lung diffusing capacity ...
Page 88: ...88 Quark PFT User Manual ...
Page 89: ...MIP MEP option ...
Page 91: ...Pulse oximetry option ...
Page 93: ...Exercise testing ...
Page 116: ...116 Quark PFT User Manual ...
Page 117: ...Resting Metabolic Rate Test ...
Page 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Page 124: ...124 Quark PFT User Manual ...
Page 125: ...Sub maximal Exercise Testing ...
Page 131: ...System maintenance ...
Page 134: ...134 Quark PFT User Manual ...
Page 137: ...Appendix ...
Page 148: ...148 Quark PFT User Manual ...