Chapter 12 - Exercise Testing - 111
Software configuration
The software can be customised as you wish. Most of the feature are easily editable to
be tailored according to different purposes.
Data viewing
The software allows to calculate a huge number of parameters; it is advisable, in order
to simplify the analysis of the results, view in the data window the only desired
parameters.
Select the parameters to view
1. Choose
Parameters to view/Test visualisation…
from the
Options
menu.
2. Select the parameters you require to view.
3. Press
OK
to confirm the selected configuration.
Select the parameters to view during the test
1. Choose
Parameters to view/Test execution…
from the
Options
menu.
2. Select the parameters you require to view.
3. Press
OK
to confirm the selected configuration.
Sort the parameters
It is possible to sort the parameters (both for viewing and printing purposes) according
to the desired order.
1. Select
Sorting parameters
from the
Options
menu.
2. Move the parameters in the order you want by pressing and holding the left mouse
button.
3. Press
OK
to maintain the current configuration.
Steady State
The program has an algorithm to tag sets of steps as Steady State.
The algorithm considers belonging to the Steady State the only consecutive steps that
meet the following conditions:
•
The value of VO2, VE and R do not vary from their mean values more than
Threshold (%)
;
•
The number of consecutive steps that met the preceding criteria are at least
Minimum number of steps
.
Customise the Steady State detection criteria
1. Choose
Steady State
from the
Options
menu
2. Type the desired values for
Minimum number of steps
,
VO
2
threshold (%)
,
VE
threshold (%)
and
R threshold
.
The steps which satisfy these conditions will be highlighted with a yellow bar.
Summary of Contents for Quark PFT
Page 11: ...Getting started ...
Page 27: ...Measurements ...
Page 32: ...32 Quark PFT User Manual ...
Page 33: ...Installation ...
Page 42: ...42 Quark PFT User Manual ...
Page 43: ...Calibration ...
Page 50: ...50 Quark PFT User Manual ...
Page 51: ...Database Management ...
Page 57: ...Spirometry ...
Page 72: ...72 Quark PFT User Manual ...
Page 73: ...Lung Volumes ...
Page 78: ...78 Quark PFT User Manual ...
Page 79: ...P0 1 Respiratory drive ...
Page 82: ...82 Quark PFT User Manual ...
Page 83: ...Lung diffusing capacity ...
Page 88: ...88 Quark PFT User Manual ...
Page 89: ...MIP MEP option ...
Page 91: ...Pulse oximetry option ...
Page 93: ...Exercise testing ...
Page 116: ...116 Quark PFT User Manual ...
Page 117: ...Resting Metabolic Rate Test ...
Page 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Page 124: ...124 Quark PFT User Manual ...
Page 125: ...Sub maximal Exercise Testing ...
Page 131: ...System maintenance ...
Page 134: ...134 Quark PFT User Manual ...
Page 137: ...Appendix ...
Page 148: ...148 Quark PFT User Manual ...