48 - Quark PFT User Manual
Calibrating the turbine
After having run the calibration program:
1. Select
Reference Values
from the
File
menu. If your syringe has a different value
from the default one (3 litres), please enter the correct value.
2.
Select
Calibration/Turbine…
.
3. When
the
Calibration Turbine
dialog box appears with the syringe piston initially
pushed all the way in, move the piston in and out for 5 inspiratory strokes and 5
expiratory strokes in order to get the first values appearing on the display. Then
move the syringe piston for other 10 strokes (IN and EX).
4. At each of the 10 steps the software displays the results of the manoeuvre and the
percentage error in the reading.
5. At the end of this operation, the software displays the new calibration factors. Press
OK
to store the new value.
Note
: if you are using a slow
PC, we recommend to set
an higher refresh time.
Summary of Contents for Quark PFT
Page 11: ...Getting started ...
Page 27: ...Measurements ...
Page 32: ...32 Quark PFT User Manual ...
Page 33: ...Installation ...
Page 42: ...42 Quark PFT User Manual ...
Page 43: ...Calibration ...
Page 50: ...50 Quark PFT User Manual ...
Page 51: ...Database Management ...
Page 57: ...Spirometry ...
Page 72: ...72 Quark PFT User Manual ...
Page 73: ...Lung Volumes ...
Page 78: ...78 Quark PFT User Manual ...
Page 79: ...P0 1 Respiratory drive ...
Page 82: ...82 Quark PFT User Manual ...
Page 83: ...Lung diffusing capacity ...
Page 88: ...88 Quark PFT User Manual ...
Page 89: ...MIP MEP option ...
Page 91: ...Pulse oximetry option ...
Page 93: ...Exercise testing ...
Page 116: ...116 Quark PFT User Manual ...
Page 117: ...Resting Metabolic Rate Test ...
Page 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Page 124: ...124 Quark PFT User Manual ...
Page 125: ...Sub maximal Exercise Testing ...
Page 131: ...System maintenance ...
Page 134: ...134 Quark PFT User Manual ...
Page 137: ...Appendix ...
Page 148: ...148 Quark PFT User Manual ...