Chapter 15 - System maintenance - 133
Cleaning the turbine flowmeter
It is necessary to disinfect periodically the turbine for sanitary measures or/and for the
correct device function.
As disinfecting solution it is suggested Sodium hypochlorite 5% (bleach).
The disinfecting procedure is easy and may be effected every time the user needs,
keeping attention to some precautions:
1. Take out the turbine.
2. Dip it in a disinfectant solution (non alcoholic based) for about 1 hour.
3. Rinse the turbine in a vessel, filled of clean water, shaking gently to remove the
disinfectant (do not clean the turbine by putting it under running water!).
4. Let it dry to air.
5. After cleaning the turbine, check if the turbine propeller rotates freely even with a
low speed air flow.
6. Connect the turbine to the reader.
Precautions during the cleaning of the turbine
1. Do not expose the turbine to high heat and do not put it under running water.
2. Do not ever dip the optoelectronic reader in any kind of solution, the liquid
infiltration would damage the internal circuit.
3. Do not use alcoholic solutions to clean the turbine.
Suggested disinfection solutions
Helipur H Plus Braun Melsungen AG
Gigasept FF
Schulke & Mayr GmbH
Dismozon pur
Bode Chemie GmbH
TETA-S Fresenius
AG
CIDEX
Johnson & Johnson
Cleaning the Breathing valve
The breathing valve should be cleaned everyday or at least every 10-15 tests.
Warning:
Do not use
alcoholic solutions for the
turbine, otherwise there can
be damages to the plastic
material.
Summary of Contents for Quark PFT
Page 11: ...Getting started ...
Page 27: ...Measurements ...
Page 32: ...32 Quark PFT User Manual ...
Page 33: ...Installation ...
Page 42: ...42 Quark PFT User Manual ...
Page 43: ...Calibration ...
Page 50: ...50 Quark PFT User Manual ...
Page 51: ...Database Management ...
Page 57: ...Spirometry ...
Page 72: ...72 Quark PFT User Manual ...
Page 73: ...Lung Volumes ...
Page 78: ...78 Quark PFT User Manual ...
Page 79: ...P0 1 Respiratory drive ...
Page 82: ...82 Quark PFT User Manual ...
Page 83: ...Lung diffusing capacity ...
Page 88: ...88 Quark PFT User Manual ...
Page 89: ...MIP MEP option ...
Page 91: ...Pulse oximetry option ...
Page 93: ...Exercise testing ...
Page 116: ...116 Quark PFT User Manual ...
Page 117: ...Resting Metabolic Rate Test ...
Page 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Page 124: ...124 Quark PFT User Manual ...
Page 125: ...Sub maximal Exercise Testing ...
Page 131: ...System maintenance ...
Page 134: ...134 Quark PFT User Manual ...
Page 137: ...Appendix ...
Page 148: ...148 Quark PFT User Manual ...