132 - Quark PFT User Manual
System maintenance
All service operations which are not specified in this user manual should be performed
by qualified personnel in accordance with the service handbook (to be required to the
manufacturer).
Rubber mouthpieces, face masks, breathing valve and the other parts are not shipped
sterile. They should be disinfected before using according to the following instructions.
All materials used in the construction of the Quark PFT are non toxic and pose no safety
risk to the patient or operator.
Prior to the device cleaning, disinfection and inspection it is necessary to switch off the
device and to disconnect adapters from the supply mains.
In order to guarantee the highest accuracy of measurements we recommend you to
disinfect the turbine periodically.
Cleaning and disinfection
Cleaning and disinfecting instructions are of fundamental importance to control
infections and assure patient safety. In fact aspiration of residue, particles and
contaminated agents are life – threatening.
In this handbook we strongly recommend you to follow the rules worked out by ATS
and ERS (see: ”Lung Volume Equipment and Infection Control” – ERS/ATS
WORKSHOP REPORT SERIES, European Respiratory Journal 1997; 10: 1928 –
1932), which are summarised and adapted for COSMED products as follows:
•
Accessible internal as well as external surfaces of equipment exposed to expired
gas should be washed and disinfected prior to testing of subsequent patients.
•
Disinfecting should ideally be performed by heat sterilisation, but gas or liquid
sterilisation can be used if the equipment is well cleaned first (no droplets of
saliva/sputum remain).
•
Disposable gloves should be worn when handling mouthpieces, when cleaning
equipment exposed to saliva or sputum and especially when drawing blood.
•
Laboratory staff should wash hands prior to testing of each patient.
•
Adopt particular precautions when testing patients with recognised high – risk
communicable diseases (e.g. tuberculosis, multidrug – resistant staphylococcus). In
these cases, the clinical need for such testing should justify the risks.
During the disinfection:
•
do not use alcohol or other liquids containing gluteraldehyde on the exterior
surfaces of the equipment. Actually they can damage polycarbonates plastics and
may produce unhealthy substances.
•
do not use abrasive powders or glass cleaners containing alcohol or ammonia on the
plexiglas components of the equipment
•
do not steam autoclave any parts of the equipment unless it is clearly specified.
•
do not immerse the optoelectronic reader.
Summary of Contents for Quark PFT
Page 11: ...Getting started ...
Page 27: ...Measurements ...
Page 32: ...32 Quark PFT User Manual ...
Page 33: ...Installation ...
Page 42: ...42 Quark PFT User Manual ...
Page 43: ...Calibration ...
Page 50: ...50 Quark PFT User Manual ...
Page 51: ...Database Management ...
Page 57: ...Spirometry ...
Page 72: ...72 Quark PFT User Manual ...
Page 73: ...Lung Volumes ...
Page 78: ...78 Quark PFT User Manual ...
Page 79: ...P0 1 Respiratory drive ...
Page 82: ...82 Quark PFT User Manual ...
Page 83: ...Lung diffusing capacity ...
Page 88: ...88 Quark PFT User Manual ...
Page 89: ...MIP MEP option ...
Page 91: ...Pulse oximetry option ...
Page 93: ...Exercise testing ...
Page 116: ...116 Quark PFT User Manual ...
Page 117: ...Resting Metabolic Rate Test ...
Page 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Page 124: ...124 Quark PFT User Manual ...
Page 125: ...Sub maximal Exercise Testing ...
Page 131: ...System maintenance ...
Page 134: ...134 Quark PFT User Manual ...
Page 137: ...Appendix ...
Page 148: ...148 Quark PFT User Manual ...