File : CS-SUP-MUTI-WTHERMAPP-E01.sxw
The precision levels may be described as follows,
Precision = 0x00 :
normal precision
(fastest measurement)
Precision = 0x01 :
high precision
Precision = 0x02 :
very high precision
Precision = 0x03 :
maximum precision
(slowest measurement)
Remark: Above all, the primary function is to preserve measurement precision, even in an
environment with considerable interference, and not to improve the measurement precision of the
probe itself.
Furthermore, the higher the precision, the higher the module power consumption. It is therefore
important to find the best possible consumption/precision compromise.
The precision level in datalogging mode is selected by configuring the parameter 0x0C.
7.4.3 Activating the datalogging mode
The datalogging mode is activated (or deactivated) by setting the bits [b3:b2] in the
Operating Mode
byte. In
standard datalogging, the 'Stop Memory Full' mode is not allowed.
➢
Operating Mode
:
MSB
LSB
Operanting Mode
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
-
Threshold
Detection Mode
0
: successive mode
1
: cumulative mode
Low Threshold
Detection
0
: deactivated
1
: activated
High Threshold
Detection
0
: deactivated
1
: activated
Datalogging
00
: deactivated
01
: time steps
10
: once a week
11
: once a month
Stop Mode of the
Datalogging
0 :
permanent loop
1 :
stop memory full
-
Attention, stopping then restarting the datalogging mode induce the re-initialization of the storage
table. In this case, the logged values will be lost.
WaveTherm modules – application handbook
page 42 of 65