About Calibration
Although this board is calibrated before shipping, you can use the calibration program to calibrate
analog input yourself.
Starting the calibration program
Click the [Calibration] button on the property page for the device to start the calibration program.
Proceed with connecting the calibration equipment and performing the calibration in accordance with
the instructions displayed by the calibration program.
Analog input calibration
Analog input calibration requires a reference voltage generator.
As the analog input has 16-bit resolution, use a reference voltage generator with a precision of at least 5
digits after the decimal point.
Calibrate one channel only for each range that you use.
Factory setting
You can use the calibration program to restore the factory calibration settings.