The "Auto-Power-Off" function can be switched off by pressing
and holding the "PEAK" button while switching on the measuring
instrument. As a result of this, all display segments appear. Press
the "PEAK" button once again. The display functions normally
again. The "et Off" symbol disappears. The measuring instrument
will now remain switched on until the battery is exhausted (empty). 

C2 Button assignments

a) "PEAK" (blue-green) 
With this button, on the one hand the colour-coded functions abo-
ve the respective switch symbol are set and on the other hand sel-
ection is made between the measurement of DC, AC and AC+DC
values during voltage and current measurement.
In addition, PEAK measurement is switched on or off. Peak measu-
rement means that alternately the maximum and minimum values
can be called via the "D-H" button.
During resistance measurement, the button selects between pure-
ly-acoustic continuity testing and actual resistance measurement.
During measurement of pulse width or when measuring the mark-
to-space ratio of square wave voltages, press the button longer
than approx. 1s. In this connection, selection can be made bet-
ween triggering on the positive "+" or negative "-" edge.
For Peak measurement, proceed as follows:
Press the Peak button for longer than approx. 1s. As a result of this
a time indicator appears in the small display. This indicates the 
time of each peak value after measurement begins. Depending on
the "DH MAX" or "DH MIN" setting, if a new higher or lower peak
value is captured, an acoustic signal sounds. The selection of "DH
MAX" or "DH MIN" is made with the "D-H" button. The "DUAL"
button resets peak measurement (resets the time indicator and 
value indicator to 0).
Peak measurement is started on the lowest measuring range with
AUTO switched off; if instead of a measured value "OL" for Over-

68

5. Mit dem Taster “DH” gelangen zur nächsten Dezimalstelle nach

links. Somit lassen sich Zeiteinstellungen von 1 s bis 99999 s
einstellen.

6. Nach erfolgter Zeiteinstellung stellen Sie den Pegelsprung nach

Ablauf des Timers ein. Mit dem Taster “DUAL” sind folgende Ein-
stellungen möglich:
Lo - Hi  /  Hi - Lo   /   Lo - PS   /   Hi - PS

7. Mit dem Taster “Hz” können Sie den “Count-Down-Timer” 

(Abwärtszähler) starten. Nach Ablauf der eingestellten Zeit erfolgt
ein kurzer Pegelsprung. Mit dem Taster “PEAK” läßt sich der 
Timer zurücksetzen.     

N  Gebrauch des Analog-Bargraphen

Der Bargraph ist leicht bedienbar und verständlich. Er ist vergleich-
bar mit dem Zeiger eines Analogmeßinstrumentes, ohne dessen
mechanische Nachteile. Er eignet sich besonders für sich schnell
verändernde Meßsignale, für welche die Digitalanzeige zu “lang-
sam” ist. So lassen sich auch Tendenzen einer Meßwertänderung
schnell erkennen und auswerten. 
Der Analogbargraph besteht aus 21 Segmenten. Er hat eine gering-
fügig höhere Meßgeschwindigkeit als die Digitalanzeige. Bei einer
Bereichsüberschreitung sind alle Segmente sichtbar.

 

Gebrauch des Multimeters in Verbindung mit einem 
Computer

Eine umfangreiche Windows-Software unter der Bezeichnung CP -
01 (Diskette  und Schnittstellenleitung) für die Verbindung/ Kommu-
nikation mit einem Computer ist optional erhältlich.

45

- with a suitable crosspoint screwdriver, unscrew the fixing 

screws on the base of the case (3). 

- Now carefully remove the cover. 

- remove the used battery from the connecting clip and 

- replace the battery with a new one of the same type.

- Having changed the battery, place the connected battery into

the battery compartment and 

- carefully close this once more.

- When closing the battery compartment take care that the

red/black connecting clip lead is not pinched. 

Do not in any event operate the measuring device
when opened! Danger to life! 
Do not leave used batteries in the unit, because
even leak-proof batteries can corrode and thereby
release chemicals, which are both damaging to
health and can destroy the battery compartment. 
Used batteries are to be treated as waste requiring
special handling and must therefore be disposed of
without harming the environment. Special collection
containers are provided for this purpose by specia-
list dealers and in scrap yards. 

B Connecting the test leads 

When taking measurements, only ever use the supplied instrument
leads or adapters/leads which can optionally be obtained for the
DMM and consequently are matched to it. Before connecting, al-
ways check the condition of the connecting plug and test probes
and check the insulation for damage. 

66

Fehler

Mögliche Ursache

keine Strommessung 

Sicherungen für die Strombereiche i.O.?

möglich

Haben die Meßleitungen einen sicheren
Kontakt in den Meßbuchsen?
Befinden sich die Meßleitungen in der A-
bzw. uA/mA-Buchse und COM?

Keine Anzeige bei 

Ist die Batterie verbraucht?

eingeschaltetem Gerät Hat das Meßgerät nach 15-Min-Nichtge-

brauch automatisch abgeschaltet?

• Beim Öffnen von Abdeckungen oder Entfernen

von Teilen, außer wenn dies von Hand möglich ist,
können spannungsführende Teile freigelegt wer-
den. Es können auch Anschlußstellen spannungs-
führend sein. Vor einem Abgleich, einer Wartung,
einer Instandsetzung oder einem Austausch von
Teilen oder Baugruppen, muß das Gerät von allen
Spannungsquellen und Meßkreisen getrennt sein,
wenn ein Öffnen des Gerätes erforderlich ist.
Wenn danach ein Abgleich, eine Wartung oder ei-
ne Reparatur am geöffneten Gerät unter Span-
nung unvermeidlich ist, darf das nur durch eine
Fachkraft geschehen, die mit den damit verbunde-
nen Gefahren bzw. den einschlägigen Vorschriften
dafür (VDE 0100, VDE-0701, VDE-0683) vertraut ist.

• Kondensatoren im Gerät können noch geladen

sein, selbst wenn das Gerät von allen Spannungs-
quellen und Meßkreisen getrennt wurde.

47

Summary of Contents for Escort 97

Page 1: ...itung entspricht dem technischen Stand bei Drucklegung Änderung in Technik und Ausstattung vorbehalten Copyright 1999 by Conrad Electronic Ned BV Printed in Germany 10 99 MZ D GB 100 Recycling Papier Chlorfrei gebleicht 100 recycling paper Bleached without chlorine Digitalmultimeter Escort 97 Seite 4 56 Digital Multimeter Escort 97 Page 57 109 Best Nr Item No 12 00 29 GB D OPERATING INSTRUCTIONS G...

Page 2: ... andere Verwendung als zuvor beschrieben führt zur Beschä digung des Meßgerätes außerdem ist dies mit Gefahren wie z B Kurzschluß Brand elektrischer Schlag etc verbunden Das ge samte Produkt darf nicht geändert bzw umgebaut werden Die Si cherheitshinweise sind unbedingt zu beachten 5 Capacitance 4 nF 2 5 6dgts 1 pF 40 nF 2 5 6dgts 10 pF 400 nF 2 0 4dgts 0 1 nF 4 uF 2 0 4dgts 1 nF 40 uF 2 0 4dgts 1...

Page 3: ...l der Sicherungen trennen Sie das Meßgerät vom Meßkreis und schalten es aus Entfernen Sie alle angeschlossenen Leitungen und Prüfspitzen Nehmen Sie einen passenden Kreuzschlitz schraubendreher zur Hand und öffnen Sie das Gehäuse vorsich tig Entnehmen Sie die defekte n Sicherung en und ersetzen Sie diese mit solchen gleichen Typs und Nennstromstärke 9 Meas range Accuracy Resolution Frequency DC vol...

Page 4: ...e interface on the multimeter is a unidirectional IR interface advantage no elec trical connection between computer and measuring instrument b The interface in the measuring instrument is activated as follows Press and hold the DUAL button while switching on the measu ring instrument at any measuring function As a result the symbol RS232 appears in the top left corner of the display Disposal If de...

Page 5: ...ahl zu verlassen betätige Sie den Ta ster RANGE länger als ca 1s Die Zweitfunktion des Tasters besteht darin daß Sie die Eingangs impedanz bei der dBm Messung einstellen können zwischen 4 und 1200 Ohm 17 J Capacitance measurement Discharge every capacitor before connecting it to the measuring instrument When short circuiting capacitors high energy discharges can occur Take care in areas where dust...

Page 6: ...ar graph are displayed the measuring range has been exceeded or the circuit being measured is interrupted 4 For the continuity testing function press the PEAK button once As a result in the left half of the display the symbol appears With continuity testing with for example wire resis tances of less than 10 Ohm an acoustic signal sounds The con tinuity testing range extends from 0 to 400 Ohm With ...

Page 7: ...C im kleinen Display Zweimal drücken Anzeige des Puls Pausenverhältnisses in im großen Spannungsanzeige AC oder DC oder AC DC im kleinen Display Dreimal drücken Pulsbreitenanzeige in ms Halbwelle im großen Spannungsanzeige AC oder DC oder AC DC im klei 25 Press twice DC current indication in mA in the large AC current indication in mA in the small display Press three times return to normal mA DC c...

Page 8: ...e nachfolgende Abbildung Der Meßwert wird im großen Display angezeigt 29 large mV voltage indication AC or DC or AC DC in the small display Press four times longer than approx 1s return to normal mV voltage measurement AC or DC or AC DC without small dis play The mV range 40 mV and 400 mV offers an input impedance of 1000 MOhm If a minus sign can be seen in front of the res pective measured value ...

Page 9: ...ll display AC Press once AC voltage indication in the large frequency indica tion in the small display Press twice dBm indication in the large AC voltage indication in the small display Press three times AC voltage indication in the large DC voltage indication in the small display Press four times return to normal AC voltage measurement without small display AC DC Press once mixed voltage indicati...

Page 10: ...en usw auf Durchgang aku stisch überprüft werden Messen Sie keine geladenen Kondensatoren da sonst durch eine mögliche Entladung Ihr Meßgerät zerstört werden kann 5 Diese akustische Anzeige bei Unterschreitung eines bestimmten Widerstandswertes wird in den anderen Bereichen wie folgt fort gesetzt ist keine Durchgangsprüfung Im Bereich 4 000 kOhm bei R 100 Ohm im Bereich 40 00 kOhm bei R 1 kOhm im ...

Page 11: ... slowly changing measu red values or when establishing small resistances while ignoring the resistance of the test leads The bar graph indicator functions during this as a so called balance indicator The function is switched on by pressing the REL button During OL Overload indication for example during resistance measure ment the function cannot be activated The two test probes must first be held ...

Page 12: ...alogbargraph besteht aus 21 Segmenten Er hat eine gering fügig höhere Meßgeschwindigkeit als die Digitalanzeige Bei einer Bereichsüberschreitung sind alle Segmente sichtbar O Gebrauch des Multimeters in Verbindung mit einem Computer Eine umfangreiche Windows Software unter der Bezeichnung CP 01 Diskette und Schnittstellenleitung für die Verbindung Kommu nikation mit einem Computer ist optional erh...

Page 13: ...ng facilities hobby and self help workshops the handling of measuring instruments must be supervised responsi bly by trained personnel Except where this can be performed manually live parts may be exposed when opening covers or removing components Connection points may also be live Before any adjustment maintenance repair or exchange of parts or assemblies requiring opening of the unit the unit mu...

Page 14: ...xed voltages AC DC to 750 VACrms VDC max True rms measurement of DC and AC currents up to 10 A max for 30 s max protected Resistance measurement up to 40 MOhm max Frequency measurement up to 200 kHz max dBm measurement from 80 79 to 81 48 dBm Measurement of temperatures from 40 C to 1372 C via an exter nal temperature sensor Capacitance measurement from approx 10 pF to approx 10000 µF Continuity c...

Page 15: ...the measuring instrument 40 C to 0 3 3dgts 1 C 1372 C 40 F to 3 0 6 F 1 F 2502 F Tolerance of the temperature sensor DP 27 measuring range from 40 C to 204 C 2 2 C 107 Dieses Gerät ist gemäß DIN 57 411 Teil 1 VDE 0411 Teil 1 Schutzmaßnahmen für elektronische Meßgeräte bzw IEC 1010 1 gebaut und geprüft und hat das Werk in sicherheitstechnisch einwandfreiem Zustand verlassen Um diesen Zustand zu erh...

Page 16: ...gts 10 mA 1 mA Load voltage voltage drop at shunt resistance value 0 04 V 100 Ohm at 400 µA 0 4 V 100Ohm at 4000 µA 0 08 V 1 Ohm at 40 mA 0 8 V 1 Ohm at 400 mA 0 084 V 0 01 Ohm at 4 A 0 21 V 0 01 Ohm at 10 A 103 Vorstellung Dieses Digitalmultimeter mit Multifunktionsanzeige ist mit mehreren Besonderheiten ausgerüstet welche manche Messung sinnvoll ergänzen Bei den Funktionen D H Data Hold MIN und ...

Page 17: ...Taster RANGE einen höheren Meßbereich einstellen b D H Mit jedem Tastendruck kurzzeitig schalten Sie die Data Hold Funktion ein oder aus Data Hold bedeutet daß der Meßwert fest gehalten wird bis die Funktion Data Hold wieder ausgeschaltet wird Dabei zeigt die Bargraphanzeige stets den augenblicklichen Meßwert an Bei der Peakmessung schalten Sie um von der Anzeige Peak DH MAX auf Peak DH MIN c MIN ...

Page 18: ...ampli tude of 3V to the centre measuring socket The Auto Power Off automatic switch off function is disabled during signal generator operation To provide a frequency variable within these limits at TTL level pro ceed as follows 95 C3 Schalter bzw Buchsenbelegung a Meßfunktionsschalter oder Drehschalter Der Meßfunktionsschalter darf während der Mes sung auf keinen Fall verstellt werden da dadurch d...

Page 19: ...is displayed then the diode is in good condition If on the contrary a voltage value is indicated then either the object to be measured is incorrectly connected or is defective Moreover in case of measuring values 100 mV an acoustic sig nal sounds During diode testing make sure that the diode and any circuit into which it may possibly be built is ab solutely voltage free All capacitors present must...

Page 20: ...an nungsmessung ohne kleines Display AC Einmal drücken Wechselspannung in mV im großen und Fre quenz im kleinen Display Zweimal drücken dBm Anzeige im großen AC Spannungsan zeige in mV im kleinen Display Dreimal drücken AC Spannungsmessung in mV im großen DC Spannungsmessung in mV im kleinen Display Viermal drücken Rückkehr zur normalen Wechselspannungs messung AC in mV ohne kleines Display AC DC ...

Page 21: ...onsuming device bad switch 83 E mA Messung Strommessung AC DC AC DC Zur Messung von Gleich oder Wechsel oder Mischströmen in mA bis max 400 mA gehen Sie wie folgt vor 1 Verbinden Sie die schwarze Meßleitung mit der COM Buchse und die rote Meßleitung mit der uA mA Buchse 2 Stellen Sie den Drehschalter auf mA 3 Verbinden Sie die Meßleitungen in Serie mit dem Meßobjekt sie he nachfolgende Abbildung D...

Page 22: ...en damit das Meßgerät nicht be schädigt wird und dadurch für Sie Lebensgefahr be stehen kann Messen Sie im A Bereich auf keinen Fall Ströme über 10 A 4 mA Messen Sie nur in Stromkreisen die selbst mit 16 A abgesichert sind G Durchgangsprüfung und Widerstandsmessung Vergewissern Sie sich daß alle zu messenden Schaltungsteile Schaltungen und Bauelemente so wie andere Meßobjekte unbedingt spannungslo...

Page 23: ...sted within the range from 4 to 1200 Ohm using the RANGE button see symbol above the button 5 The smaller display can be switched on in addition by pressing the DUAL button Press once AC voltage indication in the lar ge frequency indication in the small display Press twice dBm indication in the large voltage indication in the small display Press three times dBm indication in the large frequency in...

Page 24: ... kann Schließen Sie keine Spannungen an da dadurch das Meßgerät ebenfalls beschädigt zerstört werden kann M Timerbetrieb Das TRUE RMS Digitalmultimeter ESCORT 97 läßt sich auch als Ti mer einsetzen Nach Ablauf einer voreingestellten Zeit von 1s bis 99999 s wird ein Signal am Ausgang geschaltet Dieses Signal ist entweder ein Hi Lo Sprung negativ oder ein Lo Hi Sprung positiv oder ein positiver Puls...

Page 25: ... switch must be set to OFF 67 Wartung und Kalibrierung Um die Genauigkeit des Multimeters über einen längeren Zeitraum zu gewährleisten sollte es jährlich einmal kalibriert werden Die Ka librierung kann von unserer Kalibrierstelle im Service 2000 ko stengünstig durchgeführt werden Der Sicherungswechsel ist unter den Sicherheitsbestimmungen be schrieben Den Batteriewechsel finden Sie unter Gebrauch...

Page 26: ... used in areas or under adverse environmental conditions in which inflammable gas fu mes or dust are present or can occur For your own safety it is 63 Wechselstrom ACA TRUE RMS von 5 bis 100 des Meßbereiches 400 uA 1 0 5dgts 0 1 uA 0 01 uA 45 Hz bis 2kHz 4000 uA 1 0 5dgts 1 uA 0 1 uA 45 Hz bis 2 kHz 40 mA 1 0 5dgts 10 uA 1 uA 45 Hz bis 2 kHz 400 mA 1 0 5dgts 0 1 mA 10 uA 45 Hz bis 2 kHz 4 A 1 0 5d...

Page 27: ...rator 6 COM input socket COM or negative connection 7 V Ohm I input socket positive connection 59 Rechteckausgang Signalgenerator 0 5 1 2 10 50 60 75 100 150 200 300 600 1200 1600 2400 4800 Hz Ausgangsfrequenz Puls Pausenverhältnis einstellbar von 1 bis 99 Toleranz 0 4 Spannungshöhe Amplitude fixiert 3V 0 2V Ausgangsimpedanz Max 3 5 kOhm Maximale Eingangsgrößen Überlastschutz Spannungsmessung 1200...

Page 28: ...ls Anwender diese Bedienungsan leitung beachten Bei Fragen wenden Sie sich bitte an unsere Technische Beratung Deutschland Tel 0180 531 21 19 Mo Fr 08 00 bis 18 00 Uhr Technische Beratung Österreich Tel 0 72 42 20 30 60 Mo Do 8 30 bis 12 00 Uhr 13 00 bis 16 00 Uhr Fr 8 30 bis 12 30 Uhr Technische Beratung Schweiz Tel 0848 87 78 11 Mo Do 8 00 bis 12 00 Uhr 13 00 bis 17 00 Uhr Fr 8 00 bis 12 00 Uhr ...

Reviews: