38
UPS3000/UPS3110/UPS3210 Operation and Maintenance Manual
NOTE:
If pressure cannot be attained loosen locknut on shaft, rotate knob a few turns counter-clockwise, retighten
locknut. If you hear the pressure relief valve then rotate regulator knob counter-clockwise until relief valve shuts off.
11. Remove locknut from threaded shaft, and rotate knob counter-clockwise until bottoming out on large
locknut. After touching large locknut rotate knob clockwise 1/8 turn. Hold knob in position, install and
tighten the locknut with 40 - 50 in-lbs of torque using a 1/2" socket.
12. Open
PRESSURE LIMIT MONITOR
completely, by turning regulator knob clockwise. If you reach between
100-105% of Full Scale and pressure relief valve was not disturbed regulator has been adjusted properly.
13. Replace regulator knob cap.
4.2.14
Panel Gauge Removal
Tools required: Phillips screwdriver
7/16" Wrench
9/16 " Wrench
NOTE:
Call CONDEC for replacement Part No.’s.
1. Remove cover from its enclosure as described in Section 4.2.1 on page 28, and carefully place on a
bench top.
2. Disconnect the tubing section that connects to the gauge fitting.
3. Loosen the two thumb-nuts that hold the gauge mounting U-clamp.
4. While gripping the square portion of the gauge port with the 9/16" wrench, remove the Female Tube
Connector (PN 59721) from the Gauge.
5. Remove the two thumb-nuts, the mounting U-clamp, and the gauge.
4.2.15
Panel Gauge Installation
Tools required:
Phillips screwdriver
7/16" Wrench
9/16" Wrench
A/R 1/4" wide Teflon tape (PN 60575)
Snoop, liquid leak gas detector (PN 64781)
NOTE:
Call CONDEC for replacement Part No.’s.
1. Before installing a new gauge, wrap two layers of new Teflon tape on the port.
2. Install Gauge into panel, secure with U-clamp and tighten the two thumb screws.
3. While gripping the square portion of the gauge port with the 9/16" wrench, tighten the Female Tube
Connector (PN 59721) on to the gauge.
4. Attach the tubing section that connects to the gauge fitting.
5. Using a pressure source set input pressure to approximately 100% of Full Scale rating of UPS3110 and
check all fittings for leaks.
6. Install cover on its enclosure as described in Section 4.2.1 on page 28.
Summary of Contents for UPS3000
Page 2: ...2 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 50: ...46 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 53: ...48 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......