28
UPS3000/UPS3110/UPS3210 Operation and Maintenance Manual
4.0
Maintenance & Service
4.1
Troubleshooting
4.2
Maintenance & Service Procedures
This section outlines the mechanical and BASIC electrical repair procedures for the portable pneumatic pressure
calibrator, model UPS3000/UPS3110/UPS3210. The repair procedures cover the major components and
sub-assemblies which are critical to the proper functioning of the calibrators and that will likely need periodic
maintenance over the life of the unit. Only those persons who are formally trained as skilled technicians should
attempt to repair these units. All relevant safety precautions should be observed due to the presence of electrical
components and high-pressure.
4.2.1
Case Removal and Installation
UPS3000 External Case Removal/Installation
NOTE:
Verify pressure has been vented from system prior to case removal. Although not recommended the instrument
may be fully operated with the case removed without any potentially lethal shock hazard to operating personnel, since
accessible internal voltage is nominally 25 VDC.
1. Loosen the two thumbscrews (PN 68916) located at the bottom outermost corners of the front panel.
Screws will remain captivated to front panel but will allow chassis to slide away from rear of case.
2. Use screws to slide front panel/chassis away from case.
3. If all power and pressure connections have been removed. Gently slide panel/chassis out until back edge
of panel/chassis touches lip on front of case.
4. Tilt front of panel/chassis upward slightly and remove from case. Place assembly on a bench top.
Reverse procedure for installation.
Symptom
Problem
Remedy
Display not lit
Unit will not energize
Check fuse, check power source, if
applicable check power switch
Display slowly decreases over time
Leak in system
Check all compression and pipe fittings
with snoop, bottle of liquid leak gas
detector (PN 64781)
Display does not respond when Vernier
knob is turned
No Vernier control
Readjust isolation valves on Orion;
replace O-ring on Vernier piston
Display increases or decreases when
INPUT
(Pressure) or
VENT
valves are
closed
No pressure or vent control
Replace valve seats or O-rings in valves;
check valve needles
Unit will not stay in CAL, display shows
"O", display reads a high value @ zero
PSIG
Transducer over-pressurized
Replace transducer
Low battery indicator on display
illuminates when unit is powered
Low or no battery power
Re-charge battery, check power supply
charging voltage
No display when in battery mode after
charging
Battery will not hold charge
Replace battery
Display will not zero
Perform a ZERO/SPAN calibration
Display will shift, will not be steady
Transducer drifts or possible over
pressure
Replace transducer
Table 4-1. UPS3000/UPS3110/UPS3210 Troubleshooting
Summary of Contents for UPS3000
Page 2: ...2 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 50: ...46 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 53: ...48 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......