Operation
13
2.7
UPS3000 or UPS3210 Initial Setup Sequence
For UPS3000, refer to Figure 2-1 on page 5 and Figure 2-4 on page 6.
For UPS3210, refer to Figure 2-3 on page 6 and Figure 2-6 on page 8.
1. Connect the pressure source to the instrument via the
INPUT PRESSURE
port (16), 7/16-20, 37
o
-4 AN male
fitting provided on the rear panel. It is suggested that a Cheat Seal, PN 54854, be used between
INPUT
PRESSURE
port and pressure source fitting.
NOTE:
Valves for venting and pressure input should be installed in line with the pressure source.
2. Apply power to the instrument using AC Cord (17) or, if applicable, by pressing
POWER
switch (13) on
front panel and allow it to stabilize for at least 5 to 10 minutes.
3. Select the desired full scale pressure range via the three-position rotary
RANGE SELECT
switch (6). For
the best accuracy, the selected range must be greater than, but close as possible to, the full scale range of
the Device-Under-Test.
NOTE
:
Do not switch pressure ranges during a pressure cycle.
4. If applicable or required, select the mode of operation by momentarily depressing the
ABS/GAGE
push-button switch (2). The applicable “ABS” or “GAGE” light bar (4) will be lit to provide mode of
operation.
5. To select the desired measurement display units, depress
CONV
push-button switch (3).
NOTE:
When the PEAK HOLD or MAX/MIN option is supplied with a model having the ABS or GAGE mode, switch
selectable from front panel, the “
CONV
” button will be disabled and will not be present on the front panel. Refer to
6. If operating the unit in
GAGE
mode go to Section 2.7.1 or if operating the unit in ABS mode go to
Application of pressures greater than 1.5 times the highest pressure range of the indicator may cause
calibration errors or even permanent damage to the pressure transducer.
2.7.1
UPS3000 or UPS3210 Display of Pressure Sequence GAGE Mode
1. Vent the Input Pressure port (16) to atmosphere.
2. Momentarily depress the
ZERO
push-button switch (1). The display will indicate
0.00
.
3. The instrument is now fully configured and ready to display the applied pressure.
2.7.2
UPS3000 or UPS3210 Display of Pressure Sequence Absolute (ABS) Mode
NOTE:
If local barometric pressure is not 14.7, ABS/GAGE switch selectable units only, may need barometric offset. See
“Barometric Offset - Absolute/Gage Switch Selectable Units ONLY” on page 23.
1. If only pressure measurements greater than barometric are required, continue to Step 1.1. If pressure
measurements above and below atmospheric pressure are required go to step 2.
1.1. To apply pressure, close the customer supplied vent valve attached in line with the
INPUT PRESSURE
port
(16). Unit will display applied pressure.
2. If pressure measurements above and below atmospheric pressure are required, connect a VACUUM
PUMP in line with the
INPUT PRESSURE
port (16).
3. Close the customer supplied vent and input valve attached in line with the
INPUT PRESSURE
port (16).
4. Apply power to the vacuum pump and allow it to evacuate the system for several minutes or until the
digital display reading reaches equilibrium near Zero psia. Momentarily press the
ZERO
push-button
switch
(1)
to establish a zero reference on the display.
5. The instrument is now fully configured and ready to display the applied pressure.
#AUTION
Summary of Contents for UPS3000
Page 2: ...2 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 50: ...46 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 53: ...48 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......