Safety Techniques 

 

The high  voltage circuit of the MegaPulse  Code 360 can be shut off at any time by turning OFF the rear 
power switch. Always press TRIGGER to discharge the tester before turning OFF. 
 
The MegaPulse tester is provided with a digital 

VOLTAGE ADJUST

 knob on the front panel. This voltage 

setting should always confirm by pressing the VOLTAGE ADJUST knob before start any testing. 
 
The MegaPulse tester is provided with a 

CHARGE

 switch that is in the unarmed "Standby" setting when the 

tester is first turned ON. When the yellow 

CHARGE

 button is lit, the tester will not provide high voltage 

until the 

CHARGE

 Button and the 

TRIGGER

 Button have been pressed in order. To prevent inadvertent 

operation, the operator should be instructed not to press the 

CHARGE

 Button until the test is ready. 

 
The MegaPulse tester has been designed for one-touch operation with the right hand.  If possible, it should 
be set up to the left and in front of the equipment under test. The equipment under test should be connected 
to the MegaPulse tester and then left alone by the operator. After the operator is clear of the Tester and the 
equipment  under  test,  the  operator  should  turn  the  rear-panel  power  switch  to  ON,  press  the 

CHARGE

 

Button,  adjust  the  voltage  to  the  desired  level  (as  displayed  on  the  front  panel  meter),  then  press  the 

TRIGGER

 Button, with his right hand.  This will allow the greatest separation between the operator and the 

test being conducted. Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not 
pressed. 
 
The MegaPulse tester is designed to bleed the high voltage away after the test  has concluded. In order to 
ensure that any voltage present in the equipment being tested has been completely bled away, the operator 
should not unplug the equipment under test from the MegaPulse until the front panel meter reads a safe level 
(40 volts or less is generally considered a safe level). Pressing the TRIGGER button before disconnecting 
main power (or turning the equipment off) will ensure that the internal capacitors are discharged as much as 
possible. 

Safety Markings 

 

 

Summary of Contents for MegaPulse 360

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...n and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...e Instrument 4 AC Line Voltage Requirements 5 Fuse Replacement 5 Section 3 6 Specifications and Controls 6 Front and Rear Panel Features 7 Section 4 11 Operating Instructions 11 External Interlock 11...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...uipment under test The equipment under test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of the Tester and the equipment under test the o...

Page 9: ...the OUTPUT as a negative pulse relative to the RETURN jack The polarity switch only operates when the CHARGE LED is lit i e the output is not charged Note that the voltage meter may indicate that some...

Page 10: ...please contact our customer support department at 1 800 748 6224 Product Package for Code 360 Tester Description Part Number Code 360 MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R...

Page 11: ...e certified or approved by any recognized national test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument...

Page 12: ...adjusted by frontal VOLTAGE knob or by PC TestMinder Polarity Control Positive and Negative Alternating control available with optional computer control Voltage Display 4 Digit LED Display Voltage Met...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 1 The rear panel features of the MegaPulse are...

Page 14: ...be blinking if the Interlock Switch is open Only testers with Interlock Switch Option TRIGGER indicator will blink at when the Voltage This effect will remain on until the TRIGGER switch is pressed O...

Page 15: ...9 Figure 3 MegaPulse Code 360 Rear Panel...

Page 16: ...te source of supply Fuse holder provides access for Fuse replacement and the Power Switch is used to turn the tester ON and OFF 4 TestMinder Allow the communication between the tester and computer int...

Page 17: ...has disabled the keyboard or Voltage Knob it is possible to enable them by using the next keyboard sequence 1 Turn OFF the MegaPulse P tester 2 Press and hold the TRIGGER and NOR REV buttons 3 Turn ON...

Page 18: ...E Adjust knob to change the value of the blinking digits 5 After a few seconds without any change on the VOLTAGE Adjust knob the me Charge and Trigger a Pulse The following procedure will explain how...

Page 19: ...ual voltage is present on the main storage capacitor even when the tester is first turned ON This is due to inherent charging of the internal capacitors Pressing the TRIGGER switch will discharge the...

Page 20: ...anty period Contact the manufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the manufacturer General Maintenance To avoid contaminating the...

Page 21: ...Voltage Stop by the PC command activated it is possible to disable it using the next keyboard sequence Note Disabling Voltage Stop enables the keyboard Turn OFF the MegaPulse tester Hold in the TRIGGE...

Page 22: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: