Table 1 MegaPulse Code 360 Front Panel Features 

ITEM

 

NAME

 

FUNCTION

 

VOLTAGE

 Adjust Knob 

Adjust the digital voltage set point in the tester. 
Press the voltage knob to display the voltage set point. This setting will blink for a few seconds on the front meter. 
Turn Clockwise to increase the setting Voltage Setting Point before pressing 

CHARGE

 button 

POLARITY

 switch 

Toggles the output pulse polarity from Normal to Reverse, Normal for positive and Reverse for Negative, the pulse will appear on the Output 
jack relative to the return jack 
The polarity switch only operates when the 

CHARGE 

indicator is lit and the voltage on the display meter is less than 180V.   The polarity 

is Normal when the 

NOR

 indicator is lit and Reverse when the 

REV

 indicator is lid.   

NOR REV 

indicator 

Indicates the state of the Output Polarity switch.  

NOR

 indicates Normal (Positive) position.  

REV

 indicates Reverse (Negative) position. 

CHARGE

 switch 

Starts the charge process of the tester capacitor. 
The 

CHARGE

 indicator will turn off after the 

CHARGE

 switch is pressed and the 

TRIGGER

 indicator will turn on. 

The charge process will stop after 2 minutes if the 

TRIGGER

 button is not pressed. 

CHARGE

 indicator 

This Yellow indicator is lit to show that pressing the 

CHARGE

 switch is the next logical step in a test sequence. 

CHARGE

 indicator is lit when the tester is turn ON an after pressing 

TRIGGER

 button. 

CHARGE

 indicator will go out after pressing 

CHARGE

 button. 

CHARGE

 and 

TRIGGER

 Indicators will be blinking if the Interlock Switch is open. (Only testers with Interlock Switch Option) 

TRIGGER

 switch 

Triggers the output impulse waveform. 
The impulse waveform will appear across the output leads. 

TRIGGER

 indicator 

This Red indicator is lit to show that the tester can be triggered. 

TRIGGER

 indicator is lit for 2 minutes after the 

CHARGE

 button is pressed. 

TRIGGER

 indicator will go out after pressing 

TRIGGER

 button. 

TRIGGER

 and 

CHARGE

 Indicators will be blinking if the Interlock Switch is open (Only testers with Interlock Switch Option) 

TRIGGER

 indicator will blink at when the Voltage. This effect will remain on until the 

TRIGGER

 switch is pressed. (Only testers with 

PC Interface option) 

VOLTAGE

 meter 

Displays the output voltage set point. 
The voltage reading will increase from zero to the voltage set point when the CHARGE button is pressed. Note that the Voltage meter may 
indicate that some residual voltage is present on the main storage capacitor, even when the tester is first turned ON. This is due to inherent 
charging  of  the  internal  capacitors.  Pressing  the 

TRIGGER

  switch  will  discharge  the  capacitors.    Note  that  the  peak  amplitude  of  the 

measured output waveform is proportional to the voltage that is read of the front panel of the MegaPulse, but it will always be somewhat 
lower.  This  is  because  the  meter  on  the  MegaPulse  is  measuring  the  voltage  on  the  main  impulse  storage  capacitor.  This  voltage  will 
intentionally dissipate to some extent before reaching the output leads.  
The meter will start to flash at 

5050V

 to indicate that voltage is in the maximum limits. 

If unit includes PC Interface and the Keyboard is locked, the display will show OFF when a button is pressed. 

OUTPUT

 jack 

The impulse waveform appears on the 

OUTPUT

 jack, referenced to the 

RETURN 

jack 

RETURN

 jack 

This is the return for the impulse waveform. This jack is referenced to the chassis of the MegaPulse and is referenced to earth ground if the 
MegaPulse is properly grounded. Even though this jack is referenced to ground, it should be treated as hazardous whenever the MegaPulse 
is turned ON. 

10  

INTERNAL LOAD 

Used for waveform verification purposes, connects a 50Ω load at the Output. 

11 

ENERGY MEAS. Switch 

By pressing this switch, the front display will blink showing for a few seconds the energy measurement from the last pulse. 
Note: This function only works when the MegaPulse unit is not charging or triggering.

 

12 

E-STOP Switch 

When pressed, stops tester functions and discharges voltage. Puts tester in interlock mode until it is restored. Turn knob clockwise to 
restore it to normal operation. 

Summary of Contents for MegaPulse 360

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...n and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...e Instrument 4 AC Line Voltage Requirements 5 Fuse Replacement 5 Section 3 6 Specifications and Controls 6 Front and Rear Panel Features 7 Section 4 11 Operating Instructions 11 External Interlock 11...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...uipment under test The equipment under test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of the Tester and the equipment under test the o...

Page 9: ...the OUTPUT as a negative pulse relative to the RETURN jack The polarity switch only operates when the CHARGE LED is lit i e the output is not charged Note that the voltage meter may indicate that some...

Page 10: ...please contact our customer support department at 1 800 748 6224 Product Package for Code 360 Tester Description Part Number Code 360 MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R...

Page 11: ...e certified or approved by any recognized national test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument...

Page 12: ...adjusted by frontal VOLTAGE knob or by PC TestMinder Polarity Control Positive and Negative Alternating control available with optional computer control Voltage Display 4 Digit LED Display Voltage Met...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 1 The rear panel features of the MegaPulse are...

Page 14: ...be blinking if the Interlock Switch is open Only testers with Interlock Switch Option TRIGGER indicator will blink at when the Voltage This effect will remain on until the TRIGGER switch is pressed O...

Page 15: ...9 Figure 3 MegaPulse Code 360 Rear Panel...

Page 16: ...te source of supply Fuse holder provides access for Fuse replacement and the Power Switch is used to turn the tester ON and OFF 4 TestMinder Allow the communication between the tester and computer int...

Page 17: ...has disabled the keyboard or Voltage Knob it is possible to enable them by using the next keyboard sequence 1 Turn OFF the MegaPulse P tester 2 Press and hold the TRIGGER and NOR REV buttons 3 Turn ON...

Page 18: ...E Adjust knob to change the value of the blinking digits 5 After a few seconds without any change on the VOLTAGE Adjust knob the me Charge and Trigger a Pulse The following procedure will explain how...

Page 19: ...ual voltage is present on the main storage capacitor even when the tester is first turned ON This is due to inherent charging of the internal capacitors Pressing the TRIGGER switch will discharge the...

Page 20: ...anty period Contact the manufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the manufacturer General Maintenance To avoid contaminating the...

Page 21: ...Voltage Stop by the PC command activated it is possible to disable it using the next keyboard sequence Note Disabling Voltage Stop enables the keyboard Turn OFF the MegaPulse tester Hold in the TRIGGE...

Page 22: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: