6
Section 3
Specifications and Controls
Specifications for the MegaPulse Code 360 are listed in the table below and the component
designations are shown in
Error! Reference source not found.
.
ELECTRICAL
Output Voltage:
5000 V tolerance ±1%
Main Capacitance:
35 uF ±3% (Dry capacitor type, 2.5 million cycles life)
Main Inductance:
50 mH ±3%
Energy Output:
360J ±5% with 50Ω load at the Output.
Voltage Control:
Digital Set point adjusted by frontal VOLTAGE knob or by PC
TestMinder.
Polarity Control:
Positive and Negative. Alternating control available with optional
computer control.
Voltage Display:
4 Digit LED Display.
Voltage Meter resolution:
2V
Duty Cycle:
1 pulse every 20s.
Input Voltage:
120V~ 50/60Hz
230V~ 50/60Hz (CE for EU)
ENVIROMENTAL
Operating Temperature:
15 - 40 ºC
Relative Humidity Range:
Altitude
0 - 90% non-condensing
0 - 3000 meters
GENERAL
Dimensions:
17” wide x 10.5” high x 17” in deep
Weight:
45 lbs. approx.
Figure 1 Code 360 Output Circuit
Model
Max. Voltage
C1
R1*
R2
L1
Code 360
5000V
35µF
50Ω
47kΩ
50mH
*R1 only for waveform verification
Summary of Contents for MegaPulse 360
Page 1: ...Instruction Manual...
Page 2: ...i...
Page 4: ...iii...
Page 6: ...v...