12 

 

Polarity Pulse Selection 

  

NOTE 

If the red  Trigger light is lit or  more than 200V remains on the internal capacitor, the polarity pulse 
selection will be automatically blocked for safety reasons. 

 

1.

 

Turn the rear-panel Power Switch ON. Positive polarity always is set by default after the unit is turned ON. 

2.

 

Press the POLARITY switch button to toggle between positive and negative. 

 

Voltage Set Point Adjustment 

 

NOTE 

If the front display shows the word “Off” when trying the adjustment, it means the front Voltage knob had 
been disabled by the testminder software. See section “Front Keyboard and Voltage Knob Enable” 

  

To adjust the Voltage Set Point: 
 

1.

 

Press the VOLTAGE Adjust knob one time to enable the adjustment mode, digits for thousands and hundreds 
will start blinking for a few seconds.  

2.

 

While digits still blinking, turn up or down the VOLTAGE Adjust knob to change the value of the blinking 
digits. 

3.

 

While digits still blinking, press again the VOLTAGE Adjust knob and the tens and unit digits will blinking 
for a few seconds.  

4.

 

While digits are blinking, turn up or down the VOLTAGE Adjust knob to change the value of the blinking 
digits. 

5.

 

After a few seconds without any change on the VOLTAGE Adjust knob, the me 

 

Charge and Trigger a Pulse 

 

The following procedure will explain how to generate a high voltage pulse with the MegaPulse unit. The Megapulse 
is capable to maintain a specific voltage charge selected by the Voltage Set Point.  
 

CAUTION 

High  voltage  generated  by  the  MegaPulse  tester  is  exposed  during  this  test.  A  risk  of  shock  exists.  
Exercise care when using the MegaPulse unit. 

 

1.

 

Confirm  the  Voltage  Set  Point  by  pressing  the  VOLTAGE  Adjust  Knob,  the  setting  will  blink  for  a  few 
seconds on the front meter. 

2.

 

If a different Voltage Set Point is need it, see section “Voltage Set Point Adjustment.”  

3.

 

Set the front panel switches depending the type of test need it. Follow the front panel directions or see Figure 
3 for the electrical diagram. 

4.

 

Push the yellow CHARGE button to start charging the internal high voltage capacitor and wait until the front 
meter reaches value set on step1 or 2. Verify the red TRIGGER indicator is now lit.  

5.

 

Once the desired voltage is reached, press the red 

TRIGGER

 button to deliver the high voltage pulse (be 

sure not to touch the output and return leads when pressing the trigger switch). 

6.

 

Turn the rear-panel Power Switch OFF. 

Summary of Contents for MegaPulse 360

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...n and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...e Instrument 4 AC Line Voltage Requirements 5 Fuse Replacement 5 Section 3 6 Specifications and Controls 6 Front and Rear Panel Features 7 Section 4 11 Operating Instructions 11 External Interlock 11...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...uipment under test The equipment under test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of the Tester and the equipment under test the o...

Page 9: ...the OUTPUT as a negative pulse relative to the RETURN jack The polarity switch only operates when the CHARGE LED is lit i e the output is not charged Note that the voltage meter may indicate that some...

Page 10: ...please contact our customer support department at 1 800 748 6224 Product Package for Code 360 Tester Description Part Number Code 360 MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R...

Page 11: ...e certified or approved by any recognized national test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument...

Page 12: ...adjusted by frontal VOLTAGE knob or by PC TestMinder Polarity Control Positive and Negative Alternating control available with optional computer control Voltage Display 4 Digit LED Display Voltage Met...

Page 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 1 The rear panel features of the MegaPulse are...

Page 14: ...be blinking if the Interlock Switch is open Only testers with Interlock Switch Option TRIGGER indicator will blink at when the Voltage This effect will remain on until the TRIGGER switch is pressed O...

Page 15: ...9 Figure 3 MegaPulse Code 360 Rear Panel...

Page 16: ...te source of supply Fuse holder provides access for Fuse replacement and the Power Switch is used to turn the tester ON and OFF 4 TestMinder Allow the communication between the tester and computer int...

Page 17: ...has disabled the keyboard or Voltage Knob it is possible to enable them by using the next keyboard sequence 1 Turn OFF the MegaPulse P tester 2 Press and hold the TRIGGER and NOR REV buttons 3 Turn ON...

Page 18: ...E Adjust knob to change the value of the blinking digits 5 After a few seconds without any change on the VOLTAGE Adjust knob the me Charge and Trigger a Pulse The following procedure will explain how...

Page 19: ...ual voltage is present on the main storage capacitor even when the tester is first turned ON This is due to inherent charging of the internal capacitors Pressing the TRIGGER switch will discharge the...

Page 20: ...anty period Contact the manufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the manufacturer General Maintenance To avoid contaminating the...

Page 21: ...Voltage Stop by the PC command activated it is possible to disable it using the next keyboard sequence Note Disabling Voltage Stop enables the keyboard Turn OFF the MegaPulse tester Hold in the TRIGGE...

Page 22: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: