background image

 

Pressing  the  POLARITY  switch  on  the  front  panel  can  change  the  polarity  of  the  output 
waveform.  The polarity is Normal when the 

NOR

 indicator is lit.  In this case, the high voltage 

will  appear  on  the 

OUTPUT

  as  a  positive  pulse  relative  to  the 

RETURN

  jack.    When  the 

polarity switch is in the Reverse position (

REV

 indicator is lit), the high voltage will appear on 

the 

OUTPUT

  as  a  negative  pulse  relative  to  the 

RETURN

  jack.    The  polarity  switch  only 

operates when the 

CHARGE 

LED is lit, i.e. the output is not charged. 

 

Note  that  the  voltage  meter  may  indicate  that  some  residual  voltage  is  present  on  the  main 
storage capacitor, even when the tester is first turned ON.  This is due to inherent charging of the 
internal capacitors.  Pressing the 

TRIGGER

 switch will discharge the capacitors (be sure not to 

touch the output and return leads when pressing the trigger switch). 

 

Note that the peak amplitude of the measured output waveform is proportional to the voltage that 
is  read  on  the  front  panel  of  the  MegaPulse,  but  it  will  always  be  somewhat  lower.    This  is 
because  the  meter  on  the  MegaPulse  is  measuring  the  voltage  on  the  main  impulse  storage 
capacitor.    This  voltage  will  intentionally  dissipate  to  some  extent  before  reaching  the  output 
leads.    Therefore,  it  is  important  to  measure  the  peak  amplitude  of  the  output  waveform,  and 
adjust the output of the MegaPulse accordingly. 

 

Determination  of  Passing  and  Failing  results  can  prove  difficult.    To  obtain  the  most  accurate 
results, it is generally necessary to perform multiple impulse tests on a few different test samples 
(that  have  adequate  insulation  to  pass  the  impulse  test).    Take  note  of  the  impulse  waveshape, 
amplitude,  and  duration.    Also  note  how  much  variance  there  is  in  the  waveshape  from  test  to 
test.    Also  (if  possible),  perform  impulse  testing  on  some  test  samples  that  are  known  to  have 
inadequate  (or  damaged)  insulation.    Take  note  of  the  impulse  waveshape,  amplitude,  and 
duration, when an insulation breakdown occurs.  

 

 

Summary of Contents for MegaPulse 1.2x50-12.3PF-1

Page 1: ...MegaPulse IMPULSE TESTER 1 2x50 12 3PF 1 500ohm 0 5J Instruction Manual...

Page 2: ...design and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review...

Page 3: ......

Page 4: ...ter 9 AC Line Voltage Requirements 9 Fuse Replacement 10 Front and Rear Panel Features 10 Setting the Peak Output Voltage 14 Initial Checkout Procedure 15 Testing 16 Technical Assistance 18 Maintenanc...

Page 5: ...1...

Page 6: ...are that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especia...

Page 7: ...will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to ensure that any voltage present...

Page 8: ...amplitude of the measured output waveform is proportional to the voltage that is read on the front panel of the MegaPulse but it will always be somewhat lower This is because the meter on the MegaPul...

Page 9: ...ed in the Error Reference source not found below and the component designations are shown in Figure 1 Figure 1 General Electrical diagram Output Waveform Voltage C1 R1 R2 C2 1000 Vp 0 5 J 1 2x50 us 10...

Page 10: ...e 1 as specified however the component values of R1 and C1 may be modified to account for secondary circuit effects in order to generate the specified output waveform which is described in Figure 3 as...

Page 11: ...to half value T2 are as specified in IEC 61180 1 shown in Figure 4 Figure 4 Waveform time measurements from IEC 61180 1 Recommended test voltages and tolerance on peak voltage are as specified in IEC...

Page 12: ...8 Figure 5 Test Voltages and Tolerance from IEC 60255 5...

Page 13: ...ase use the original shipping container If the original shipping container is not available be sure that adequate protection is provided to prevent damage during shipment We recommend that the instrum...

Page 14: ...ug the instrument from the source of supply 3 Remove the power inlet cord from the instrument 4 Using a small screwdriver pry open the fuse holder door 5 Replace the fuse with a new one of the correct...

Page 15: ...11 Figure 6 Controls Indicators Connectors MegaPulse 1 2x50 12 3PF 1 500ohm 0 5J Front Panel image for reference only...

Page 16: ...pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only testers with Interlock Switch Option TRIGGER indicator will blink at when the Voltage This...

Page 17: ...h is used to turn the tester ON and OFF 2 RS 232 Interface Optional Allow the communication between the tester and computer interface a RS 232 to USB is available 3 Interlock Switch Optional Emergency...

Page 18: ...OT SET THE VOLTAGE ABOVE THE MAXIMUM VOLTAGE SETTING SPECIFIED IN Table 4 DOING SO MAY DAMAGE THE EQUIPMENT Output Switch Setting Desired Peak Output Voltage Vp 1 Output Voltage Setting Vs Maximum Vol...

Page 19: ...n Toggle the POLARITY switch if necessary so that the NOR indicator is lit 5 Note that the Voltage meter may indicate that some residual voltage is present on the main storage capacitor even when the...

Page 20: ...ster on 5 Note that the Voltage meter may indicate that some residual voltage is present on the main storage capacitor even when the tester is first turned ON This is due to inherent charging of the i...

Page 21: ...Note that the peak amplitude of the measured output waveform is proportional to the voltage that is read of the front panel of the MegaPulse but it will always be somewhat lower This is because the me...

Page 22: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 23: ...will be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and...

Page 24: ...on procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse 1 2x50 12 3PF 1 500ohm 0 5J tes...

Page 25: ...the order of 10 20ms This is because the COM contact moves at a finite speed over a large contact gap distance Therefore it is reasonable to expect that the power source that is charging the capacitor...

Page 26: ...current flowing which allows the arc to be maintained Figure A 1 shows a test circuit schematic Figure A 2 shows the ideal simulated output waveform and Figure A 3 shows the actual output highlightin...

Page 27: ...s a shorter time constant Rc R2 x C3 As soon as the arc is extinguished a high impedance exists again between the COM contact and the NO contact Therefore a voltage difference will again develop betwe...

Page 28: ...lation In fact the waveform definitions from IEC 60060 1 and other standards do not specify the wave shape after the voltage has decayed to below 50 of the peak value In practice these transients have...

Page 29: ...25 Figure A 5 Breakdown as the voltage is rising...

Page 30: ...26 Figure A 6 Dieletric breakdown immediately after Vpeak...

Page 31: ...27 Figure A 7 Dielectric breakdown after Vpeak delayed breakdown...

Reviews: