9
Table 6 Between Leads (Nonisolated)
Jack/Switch
Position
FEMALE Cordset
To EUT
MALE Cordset
To Mains Power
RED LEAKAGE Jack
To voltage meter
and EUT
nonisolated patient
connections
GREEN LEAKAGE Jack
To voltage meter
RED RESISTANCE jack
N/C
FLOATING MEAS. PT.
EUT nonisolated
patient connections
MEAS. POINT Switch
FLOATING
EARTH Switch
TEST IN BOTH
POSITIONS
NEUTRAL Switch
CONNECTED
POLARITY Switch
TEST IN BOTH
POSITIONS
EUT Power Switch
ON
PASS Result:
<10uA ground
connected
<50uA ground
open
FIGURE 10.2.13.5.4.1 Test Circuit for Measuring
Leakage Current Between Patient Leads
(Nonisolated and Isolated).
Summary of Contents for LCT-601