Section 1
Introduction
The LCT-601 is designed to facilitate connections as noted in NFPA 99 so tests by manufacturers
and users can be made in a fast and repeatable, safe way. Operation Instructions are broken up so
manufacturers and end users can go directly to their battery of tests.
This Section contains a Checkout Procedure which can be used to check proper switch operation.
Section 2 describes use of the LCT-601 by Manufacturers, to verify compliance with the
requirements of NFPA 99, Section 10-2.13.
Section 3 describes use of the LCT-601 for those requirements for Installed Equipment, both cord
connected and permanently connected, in accordance with NFPA 99, Section 8.4.1.3.
Section 5 contains annual calibration instructions which can also be used for impromptu checks of
proper operation of the LCT-601.
Checkout Procedure
In order to verify proper operation of the switches on the LCT-601, this procedure may be used on a periodic
basis. In order to facilitate recordkeeping, the procedure is presented on the next page for easier copying if
desired.