56
Product Specifications
The InPen™ System is intended for use by patients at
home and in healthcare facilities.
Operating Conditions
5
°
C (41
°
F) to 40
°
C (104
°
F)
15 % to 93 % RH (non condensing)
Operating Altitude
-381 m (-1253 ft) to 3010 m (9878 ft)
Operating Pressure
80 kPa – 105 kPa
Storage Conditions
-5
°
C (23
°
F) to 40
°
C (104
°
F)
10 % to 98 % RH (non condensing)
Dimensions
16.5 cm (6.5”) x ø1.9 cm (0.8”)
Weight
35 g
Power Supply
Li-Mn battery (not replaceable)
Lifetime
1 year from activation
Accuracy
Meets limits for dose accuracy according to ISO 11608-1
Moisture Protection
IP22: Protection against insertion of large objects and
dripping water
Protection against Electrical Shock
Type BF applied part