Heating Cycle List
The following section will explain how to create a new heating cycle, and how to edit or
remove an existing heating cycle. The list contains all of the available heating cycles.
Create a new heating cycle
.
Selecting the New button located below the Heating
Cycle List will bring up a window that allows the user to create and name a new cycle.
Delete a heating cycle
.
Selecting Delete will permanently delete the selected heating
cycle. Once deleted, a cycle is only recoverable if it was backed up prior.
Enabling Mirage™.
The heater inside the sensor is part of a “temperature
compensation” system. Activating the heater requires that each component in this
system be made fully operational. In order to activate the heater, the user must first
enable the Mirage™. To do so, first click Settings on the Screen Selection tool bar and
then the Relay Control tab. Select Mirage™ under Relay 2 in the Relay Control area.
The Mirage will now respond to commands issued from the Heater Control screen.
Enable Mirage™
Selecting correct sensor.
The Temperature Sensor selector is located under the Heater tab on the Settings
screen.
Colnatec currently offers two different sensor head designs. The design being used
determines which temperature sensor setting should be selected.
Selecting the
incorrect heater sensor will cause the heater to operate incorrectly.
In order to
ensure proper heater function, use the following settings:
• Screw Cap Design
è
Select TC
• Twist Lock Design
è
Select RTD
27
Chapter 4
Heater Control