Technical Manual
CODEL
22
6. Maintenance
The Tunnel Tech 700 Series system is designed for low maintenance. The sensors utilise the latest
microprocessor technology. The AQM sensor uses an infrared source that has a life expectancy of
over 3 years.
6.1. Routine Maintenance
6.1.1. VIS
6.1.1.1. Mechanical Check
Every 6 months - clean the optical surfaces of the AQM sensor and the reflector unit.
Access to the lenses is obtained by removing the sight tube on the front of the sensor mounting
bracket; this is done by removing the retaining bolts securing the tube to the flange. Removal of the
tube reveals the two lenses. Clean the lenses with a
soft dry cloth
and replace the sight tube. There
should be no need to re-align the sensor.
Access to the reflector is gained by removing the sight tube on the front of the reflector mounting
bracket. This is done by removing the retaining bolts securing the tube to the flange to reveal the
mirror on the inside of the cover plate. Clean the mirror with a
soft dry cloth
and replace the sight
tube. There should be no need to re-align the reflector.
The next stage of the mechanical check is to check detector levels. By following the steps below,
observe the detector levels to see if they correspond with the required levels of 10000 +/- 2000.
6.1.1.2. Press
to return to the main display.
From the main display press
to reach Diagnostics. Press
to access
Diagnostic data.
.
6.1.1.3. Use the
key to reach Vis Diag, and press
to access.
6.1.1.4. Use the
to reach Vis DRx, and press
to access.
6.1.1.5.
Vis DRx:
Observe DRx Smoothed value while adjusting the optical alignment of the
instrument until this value has reached a plateau.
1. Meas - 09:24:29
CO: 21 ppm
3. DIAGNOSTICS
3.5. VIS Diag
VIS DRx
10168.3
3. DIAGNOSTICS
Summary of Contents for Tunnel Tech 700 Series
Page 2: ...Issue A Rev 0 Date 25 04 17 Ref 100303...
Page 3: ...CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Page 5: ...CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Page 6: ......
Page 10: ...Technical Manual CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Page 21: ...Technical Manual CODEL 11 Figure 7 Electrical Schematic...
Page 23: ...Technical Manual CODEL 13 Power Control Display PCB Electrochemical Cell PCB Communications...
Page 24: ...Technical Manual CODEL 14 Visibility PCB Ma Output PCB S Top mA RLY 4 5 Bottom mA RLY 1 2...
Page 37: ...Technical Manual CODEL 27 Alignment Laser Please Call CODEL mA Relay Card 803 604D...
Page 79: ......
Page 80: ......
Page 81: ...Technical Manual Page 1 CODEL...