Technical Manual
CODEL
20
5.3.2.1. Press
to return to the main display.
From the main display press
to reach Diagnostics. Press
to access Diagnostic
data.
.
5.3.2.2. Use the
key to reach E-C CH1 Diag, and press
to access.
5.3.2.3. Use the
to reach DRx, and press
to access.
5.3.2.4. By following this path, the data displayed on the screen represents the detector level
reading.
5.3.2.5. With the detector levels of E-Cell 1, 2, and 3 established, the final step of commissioning
for the electrochemical cell is the calibration, found in the following section.
1. Meas - 09:24:29
CO: 21 ppm
3. DIAGNOSTICS
3.2. E-C CH1 Diag
DRx
26200
3. DIAGNOSTICS