Section 2
General Description
Issue: A Revision: 1 Date: 24/07/19 Ref: 100606
Page: 25
The measurement of CO is calculated from a parameter Y, where :
Y
=
G - K.D2/D1
and
D2
=
the live detector output
D1
=
the reference detector output
K
=
a composite gain factor which takes account of all
opticaland electronic gains
G
=
scaling factor
Figure 6 shows the parameter Y against the CO concentration.
This is the calibration curve for the instrument and is opposite in shape to the
transmissivity curve shown in Figure 5. Each is fixed by the characteristics of the
4.7µm filter and cannot change. CODEL CO analysers make full use of this scale
shape to provide an easily attainable calibration point.
Calibration
2.2.5.5
Figure 6 : Calibration Curve