Section 6
Maintenance
Issue: A Revision: 1 Date: 24/07/19 Ref: 100606
Page: 127
+5 and 0V…………..
Supply rails for the micro-processor. DC
voltage.
12V…………………..
Supply rail for the plant status input.
+15V, -15V & 0VA….
Supply rails for the analogue current output.
VIN……………………
Output from the D/A converter. 0 to 2.5V
represents 0 to 20mA (4mA at 0.5V)
-15V1, 0V & +15V1…
Isolated supply for the analogue current
output.
0VB and +12VB……
Isolated supply for the plant status input.
F………………………
Reference wave from the transmitter unit (via
the power supply). Square wave, 30 to 45Hz,
5V pk-pk, centred on 2.5V.
0V and +12V………..
Power supply for the instrument
0V……………………
0V for the receiver.
T1……………………
Detector output without conditioning.
T2……………………
Detector output after first stage of gain.
T3……………………
Detector output after both stages of gain.
S-……………………..
0V supply to the heater cartridge.
S+…………………….
12V supply to the heater cartridge.
M+ and M-…………..
Supply to the chopper motor (+1V DC).
Power Supply
6.5.3.2
Receiver
6.5.3.3
Transmitter
6.5.3.4