Page: 122
EnergyTech 401
CODEL
Operating and Maintenance Manual
Monitoring Solutions
6.4.3
Operational Problems
Symptom
Possible Cause
Test
Result
Action
5. Output reading
permanently zero or
full scale
Incorrect
normalising
parameter setting
Enter parameter
mode and ensure
all parameters are
as required
Parameters
correct
Proceed to next possible
cause
Parameters
incorrect
Enter correct parameters
Incorrect calibration (Set
Cal value) zero gas level
Enter diagnostic
mode
and
observe the Y (60)
figure
1800 <Y <2010
Instrument OK – true level of
gas
in
duct
(output
permanently zero)
Y <1800
Proceed to next possible
cause
Detector levels low or
saturated
Enter diagnostic
mode and check
the
detector
levels
and
saturation count
Detector
levels
OK
Proceed to next possible
cause
Detector
levels
out of tolerance
Re-commission the instrument
and consult the fault-finding
table again
Output circuitry failure
Enter
set-up
mode (configure
output & attempt
to set zero &
span
Output responds
correctly
Re-commission the instrument
and consult the fault-finding
table again
Output does not
respond correctly
Replace signal processor
Symptom
Possible Cause
Test
Result
Action
6.
Averages
not
updated
Fault condition exists
Enter diagnostic
mode and check
fault condition
Fault
condition
exists
See Fault Condition
No fault condition
Proceed to next possible
cause
Plant status contact in
signal processor has
been made
Check plant status
input
LED
in
signal processor
LED off
Operation correct – stable
gas level
LED on
Plant status in use –
instrument will not update
until plant on
Summary of Contents for Energy Tech 401
Page 71: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 71 ...
Page 99: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 99 ...
Page 101: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 101 ...
Page 118: ...Page 118 EnergyTech 401 CODEL Operating and Maintenance Manual Monitoring Solutions ...