4-Operation
– Intermittent Infusion
50
NOTE:
During programming, you may backspace by using
:
Pressing
once clears the last entered digit.
Pressing
again clears the next number.
Pressing
when the value is empty displays the original value or the previous screen.
Resume Infusion, Start New Bag, or Reprogram Infusion
When an Intermittent infusion is stopped and the pump is re-started, the operator has the following
options:
(1)
Resume:
Resume the original infusion exactly where it left off
(2)
New Bag:
Resume the original infusion with a new bag
(3)
Repeat:
Repeat or re-program the infusion
1.
Intermittent Stop Screen
Press
or
to scroll to the appropriate selection
and press
.
2.
Resume
– New Bag – Reprogram Screen
Display shows a summary of the current infusion to be
started. User can scroll through the details by pressing
or
. Press
to start infusion.
=<
R
E
S
Resume
New Bag
Repeat
Press OK key
Dose Rate 100 ml/h
Dose Vol. 100 ml
Pause 5:00
Press OK to start
I
N
T
R
R