25
The pump display will show the accumulation screen where
all set data is shown, check the setting, if correct, press
to start operation.
7.
Wait untill pump displays the
“low battery” message
8.
Press
to turn the pump off.
9.
Connect the pump to a charger.
10.
Verify that the red LED indicating an empty battery lights on (lower left side of the charger above
channel one).
11.
Verify that the full battery LED lights red after 4-6 hours of charging (upper left side of the charger
above channel one).
Charger Indicator Test
1.
Click the pump into the charger and connect the charger to mains. Disconnect the power cord
from the AC power outlet.
2.
Verify that the CHARGE LED indicator (in the charger right side above channel II) is off.
3.
Connect the power cord to the AC power outlet. Verify that the CHARGE LED indicator is on.
Administration Sets
The
BodyGuard 121 Twins
infusion System should only be operated with Infusion sets approved for use
by the manufacturer and distributor.
Warning:
Do not operate the
BodyGuard 121 Twins
Infusion system with any
administration set, other than that recommended by the manufacturer or its distributor. The
optional
BodySet/MicroSet
Infusion Sets are equipped with a free flow protection valve, to
prevent a gravitational free flow hazard. Using a different set may essentially change the
administration rate and expose the patient to free-flow.
Low Battery
Connect To Charger
Rate 500 ml/h
Volume left 20 ml
Infused 0 ml
Press Ok to Start