4-Operation
– Intermittent Infusion
49
NOTE:
If pump is in
Delay Mode
, enter Start Time (24 Hour Clock) and Start Date. For further
information regarding the Delay Option, refer to Delay Program section.
9.
Press
to Start Infusion. Delay time option is
disablde, screen will show start time as “Immediate”.
10.
During program operation, the pump displays the rate in
which the program is operating.
NOTE:
If one of the values is changed during the programming mode, the pump will not recall
the new value unless changes are confirmed by pressing
.
NOTE
: If prime procedure is activated during data setting, the pump will return to the original
(previous) screen once priming has completed.
NOTE:
Whenever required to stop infusion, press
. Stop displays on the screen. Stop
mode is limited to two minutes. After two minutes an audible alarm sounds. Press
again to mute the alarm or press STOP/NO until the pump powers off.
Rate Adjustment
It is necessary to stop infusion in order to change any or all parameters of the current infusion program
To change the rate while in Stop mode:
1.
Press any number on the keypad and then press
.
2.
Adjust the rate and press
.
3.
Adjust all other parameters and press
to restart infusion.
Next
Dose
Start
Time:
Immediate
Press OK to start
Rate
25
ml/h