26
Set Based Anti Free-Flow Valve
The Anti free flow valve, BodyValve, enhances pump functioning by:
Preventing free-flow in the event the set is detached from the pump.
Preventing back-flow (reflux) in the event several infusion pumps are connected simultaneously
to the same patient.
Preventing free-flow in the event of a mechanical malfunction.
Preventing pump operation if the set has been loaded incorrectly.
Warning:
Replace the administration set every 72 hours to lessen the incidence of bacteria
formation.
Warning:
Disposables must be compatible with the medicine delivered.
Warning:
Patient or operator injury may result if package is opened or damaged, or if
damaged components are used. Visually inspect contents and package before each use.
Warning:
Air embolism can cause serious injury or death to the patient. Do not connect a
patient to the BodyGuard Infusion Pump until all trapped air has been cleared from the fluid
path. Carefully read the instruction for loading an administration set. Punctured set
components may cause air embolism as well.
Warning:
Drugs must not be administered to the epidural space unless the drugs are
specifically indicated for this purpose and are administered in accordance with the indications
included in the manufacture’s package insert accompanying the drugs. Epidural administration
of drugs other than those indicated for epidural use could result in serious injury to the patient.
Warning:
For epidural administration of drugs use
MicroSet
only.
Caution:
Component damage may occur if not installed properly. Assure all connections
are secure; do not over tighten. This will help minimize leaks, disconnection and component
damage.