4-Operation
– Intermittent Infusion
44
5.
Scroll to select
“delay before start” and press
.
6.
Turn delay option ON by pressing
or
.
7.
Press
and the screen displays Restart Pump
8.
Press
to continue.
Setting delay befor start
1.
After setting the program data pump enables setting the
delay (if applicable).
2.
Use numeric kepad to set delay time (up to 20 hours).
NOTE:
pressing
without setting delay time will start infusion immediately.
3.
During delay time, pump will operate in KVO mode. Time left will count down until 00:00.
NOTE:
After delay time infusion will start automatically.
Dose Program
The
Dose
program is the second program in the BodyGuard software, behind the
Continuous
program.
Programming the pump using ml/hr in Continuous mode is a standard default setting, yet an operator is
able to quickly select the
Dose
program in the event that this mode is desired.
The ability to program and adjust medication administration based on dose (
mcg/kg/min
), and not only
flow rate, provides flexibility for the medical staff and a higher level of safety and care for the patient.
Operating the Dose Program
1.
After power on and Self Test the following screen is appears:
Time option
>
Delay before start
Language
Volume 456 ml
Press OK for delay
C
Continuous
O Program
N Press No to Change
T