CMS-4 IN-HELMET CARBON MONOXIDE MONITOR
Page 6
© 2019 CLEMCO INDUSTRIES CORP.
www.clemcoindustries.com
Manual No. 29872
monitor when doffing (taking off) the respirator at
the end of the shift and to do any service or
calibration, including turning the monitor on and off.
Do this to make sure the monitor is fully functional before
placing it inside the helmet.
WARNING
Remove the monitor at the end of the shift, turn
it off, and store it in a clean environment. At the
beginning of each shift, turn on the monitor and
make sure it is fully functional before installing it
in the helmet. Failure to do so could result in CO
poisoning and death.
2.1.3 Installing Monitor in Apollo 60 and Apollo
600 Respirators
Refer to Section 2.1.2 for mounting the monitor in Apollo 20
respirators.
2.1.3.1
If not already done, adhere the Velcro
®
loop to
CMS-4 mounting clip, per Section 2.1.1.
2.1.3.2
Place the transparent mounting strip on the
inside of the inner lens, and into the same window
gasket groove as the inner lens, as shown in Figure 6.
Position it so it is within peripheral vision toward the side
of the window opening.
Figure 6
2.1.3.3
Remove backing from the Velcro
®
hook tape
and adhere it to the mounting strip (the loop tape
adheres to the mounting clip on the monitor) making sure
the LED lights are visible from inside the helmet.
2.1.3.4
Attach the monitor to the strip, positioned so the
buttons face toward the outer edge of the window
opening and LED lights toward the center of the window
opening. Don (put on) the helmet to make sure the
monitor LED lights are visible from inside the helmet.
2.1.3.4
Remove the monitor and prepare it for operation.
2.1.3.6
After the initial setup is done, remove and
reattach the monitor from inside the helmet.
Remove
the monitor when doffing the respirator at the end of
the shift and to do any service or calibration,
including turning the monitor on and off.
Do this to
make sure the monitor is fully functional before placing it
inside the helmet.
WARNING
Remove the monitor at the end of the shift, turn
it off, and store it in a clean environment. At the
beginning of each shift, turn on the monitor and
make sure it is fully functional before installing it
in the helmet. Failure to do so could result in CO
poisoning and death.
2.2
Turning ON and Startup Procedure
This section explains how to startup the CMS-4 and to
prepare it for operation.
2.2.1
Press and hold the POWER/MODE button until the
alarm beeps.
2.2.2
The monitor goes though the warm-up
sequence: all LCD segments turn on, backlight turns on,
and instrument beeps and vibrates. It then cycles
through calibration status, date/time, remaining battery
life, target gas, and alarm setpoint screens.
2.2.3
After the warm-up sequence, the monitor goes
into Measuring Mode (normal operating mode). It beeps,
the LCD screen (as shown in Figure 8) shows CO,
remaining battery life, a pulsing heart symbol (indicating
the monitor is in normal operating status), CO
concentrations (ppm), and the date and time (when set
per Section 2.3).
WARNING
If the monitor is in low battery warning, change
the battery as soon as possible. Do not use the
respirator if the dead battery warning is alarmed.
The monitor does not detect carbon monoxide
when it is in dead battery alarm, which can result
in death if carbon monoxide is present.
2.3
Setting the Date and Time ‒ Figure 7
NOTE: When using the CMS-4 for the first time, check
the current time and verify that it is correct for your time
zone. If it is not, set the time as follows:
As seen from inside the hood,
the strip is inserted into the
groove in the window gasket.
Attach Velcro
®
hook tape to
the inside of the strip. Make
sure the monitor’s LED lights
are visible when the helmet is
donned.