background image

 

CMS-4 IN-HELMET CARBON MONOXIDE MONITOR 

Page 1 

 

 

 

 

 

© 2019 CLEMCO INDUSTRIES CORP.  

  www.clemcoindustries.com  

  Manual No. 29872  

1.0 

INTRODUCTION 

 

1.1 

Scope of Manual

 

 

1.1.1

  These instructions cover operation, maintenance, 

troubleshooting  and  replacement  parts  for  the  CMS-4 
Respirator-Mounted Carbon Monoxide Monitor.  

 
1.1.2

 

The  monitor  detects  the  presence  of  carbon 

monoxide  (CO)  inside  Apollo  supplied-air  respirators, 
where  the  maximum  CO  exposure  limit  in  the  United 
States is 10 parts per million (ppm). This is the limit set 
to meet the requirement for Grade D quality breathing air. 
The  monitor  also  displays  the  PEAK  level  of  CO,  short-
term  exposure  limit  (STEL),  and  time-weighted  average 
(TWA).  Instructions  to  toggle  through  the  PEAK  CO 
concentrations,  STEL,  and  TWA  are  explained  in 
Section 4.3.

 

 

1.1.3

  All  respirator  users  and  those  responsible  for 

maintenance  and  calibration  of  the  monitor  must  read 
and  understand  this  manual  before  using  the  respirator 
or operating the monitor.

 

 

 
 
1.2 

Safety Alerts 

 

1.2.1 

Clemco uses safety alert signal words, based on 

ANSI  Z535.4-2011,  to  alert  the  user  of  a  potentially 
hazardous  situation  that  may  be  encountered  while 
operating this equipment. ANSI's definitions of the signal 
words are as follows: 
 

 

 

This is the safety alert symbol. It is 
used to alert you to potential physical 
injury hazards. Obey all safety 
messages that follow this symbol to 
avoid possible injury or death. 

 
 

 

NOTICE

 

Notice indicates information that is considered 
important, but not hazard-related, if not 
avoided, could result in property damage. 

 

 

 

 

CAUTION 

Caution indicates a hazardous situation that, if 
not avoided, could result in minor or moderate 
injury. 

 

 

WARNING 

Warning indicates a hazardous situation that, if 
not avoided, could result in death or serious 
injury. 

 

 

 

DANGER 

Danger indicates a hazardous situation that, if 
not avoided, will result in death or serious 
injury. 

 

 

1.3 

Table of Contents 

 

Topic 

Page  Section 

 

INTRODUCTION  ..........................................1 ........... 1.0 

Scope of Manual  ...........................................1 ........... 1.1 
Safety Alerts  .................................................1 ........... 1.2 
Table of Contents  .........................................1 ........... 1.3 
Specifications  .............................................. 2  .........  1.4 
Description of Operation  .............................. 2  .........  1.5 
Ancillary Equipment Requirements  ............. 3  .........  1.6 
Components and Functions

 

 ......................... 3  .......... 1.7 

  Case  ......................................................... 3  ......  1.7.1 
  Sensor Retainer  ....................................... 3  ......  1.7.2 
  Sensor  ...................................................... 3  ......  1.7.3 
  Sensor Gasket ........................................... 3  ......  1.7.4 
  Charcoal Filter (black)  .............................. 3  ......  1.7.5 
  Hydrophobic Filter (white)  .........................3  ....... 1.7.6  
  LCD (liquid crystal display) ........................ 3  ......  1.7.7 
  Control Buttons  ......................................... 4  ......  1.7.8 
  Alarm-Lights  ............................................. 5  ......  1.7.9 
  Audible Alarm  ........................................... 5  ....  1.7.10 
  Vibrating Alarm  ......................................... 5  ....  1.7.11 
  Lithium Battery (a spare is included) ......... 5  ....  1.7.12 

 

SETUP - INSTALLATION, STARTUP, 
AND OPERATION  .......................................5 ........... 2.0

 

Installation  .....................................................5 ........... 2.1 

Prepare Mounting Clip  ...............................5 ........ 2.1.1 
Installation in Apollo 20  .............................5 ........ 2.1.2 
Installation in Apollo 60 and Apollo 600  ....6 ........ 2.1.3 

Turning ON and Startup Procedure  ..............6  .......... 2.2 
Setting Date and Time  ..................................6 ........... 2.3 
Performing a Fresh-Air Adjustment  ..............7  .......... 2.4 
Turning OFF the CMS-4  ...............................7  .......... 2.5 

 

ALARMS  ......................................................8 ........... 3.0 

Alarm Indications  ..........................................8 ........... 3.1 
Alarm Indications Table  ................................8 ........... 3.2 
Resetting CO Alarms  ....................................8 ........... 3.3 
Responding to CO Alarms  ............................8 ........... 3.4 

CO Alarm  ...................................................8 ........ 3.4.1 
Over Range Alarm  .....................................8 ........ 3.4.2 
Battery Alarms  ...........................................8 ........ 3.4.3 

Summary of Contents for CMS-4

Page 1: ...om DATE OF ISSUE 07 2019 REVISION Do not use this equipment before READING this MANUAL and UNDERSTANDING its contents These WARNINGS are included for the health and safety of the operator and those in the immediate vicinity Failure to read and understand these warnings can result in injury or death Electronic files include a preface containing the same important information as in the orange cover ...

Page 2: ...ast equipment is only one component of an abrasive blasting job Other products such as air compressors air filters and receivers abrasives scaffolding hydraulic work platforms or booms equipment for lighting painting ventilating dehumidifying parts handling or specialized respirators or other equipment even if offered by Clemco may have been manufactured or supplied by others The information Clemc...

Page 3: ... other non breathable gas Never modify or change respirator air line connections without first testing the content of the line for safe breathing air Failure to test the line may result in death to the respirator user Breathing air quality must be at least Grade D as defined by the Compressed Gas Association specification G 7 1 per OSHA Regulation 29 CFR 1910 134 When compressed air is the breathi...

Page 4: ...terval clean all filters screens and alarm systems If spring loaded abrasive valves are used always cage spring before disassembly WARRANTY The following is in lieu of all warranties express implied or statutory and in no event shall seller or its agents successors nominees or assignees or either be liable for special or consequential damage arising out of a breach of warranty This warranty does n...

Page 5: ...ens valves for tightness leaks and wear Always ground the machine to eliminate hazard of static shock Install a blast machine screen to keep out foreign objects Use a blast machine cover if left outdoors overnight Never exceed the maximum working pressure of the vessel 7 AIR LINE FILTER moisture separator installed as close as possible to the blast machine inlet and sized to match the size of the ...

Page 6: ...ol to avoid possible injury or death NOTICE Notice indicates information that is considered important but not hazard related if not avoided could result in property damage CAUTION Caution indicates a hazardous situation that if not avoided could result in minor or moderate injury WARNING Warning indicates a hazardous situation that if not avoided could result in death or serious injury DANGER Dang...

Page 7: ...ion of Operation WARNING The CMS 4 detects carbon monoxide which can be life threatening When using the CMS 4 follow the instructions maintenance calibration schedule and warnings in this manual to assure proper and safe operation of the monitor and to minimize the risk of personal injury Carbon monoxide poisoning can result in death or serious injury 1 5 1 The CMS 4 respirator mounted carbon mono...

Page 8: ...ochemical cell that consists of two precious metal electrodes in a dilute acid electrolyte A gas permeable membrane covers the sensor face and allows gas to diffuse into the electrolyte The gas reacts in the sensor and produces a current proportional to the concentration of the target gas The current is amplified by the CMS 4 s circuitry converted to a measurement of gas concentration and displaye...

Page 9: ...ress and hold AIR b Immediately press POWER MODE and then release both buttons NOTE Each screen displays for 20 seconds If you do not press a button within 20 seconds the CMS 4 automatically returns to Measuring Mode Button Function POWER MODE Turns the monitor ON and OFF Activates Display Mode Enters instruction into CMS 4 s microprocessor Resets alarm after an alarm condition Also Turns the LCD ...

Page 10: ...RATION 2 1 Installation This section explains the initial installation of the CMS 4 to the inside of the respirator 2 1 1 Prepare Mounting Clip 2 1 1 1 Adhere Velcro loop to CMS 4 mounting clip Trim the Velcro loop strip to 3 4 x 1 Remove the backing and adhere it to the bottom of the monitors mounting clip as shown in Figure 4 2 1 2 Installing Monitor in Apollo 20 Respirator Refer to Section 2 1 ...

Page 11: ... the monitor when doffing the respirator at the end of the shift and to do any service or calibration including turning the monitor on and off Do this to make sure the monitor is fully functional before placing it inside the helmet WARNING Remove the monitor at the end of the shift turn it off and store it in a clean environment At the beginning of each shift turn on the monitor and make sure it i...

Page 12: ...air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 4 2 Turn on the monitor by pressing and holding the POWER MODE button until the alarm beeps 2 4 3 The monitor goes though the warm up sequence as noted in Paragraph 2 2 2 2 4 4 After the warm up sequence the monitor goes into Measuring Mode normal operating mode it beeps and the LCD displa...

Page 13: ...d to removing the respirator as soon as it is safe to do so 3 4 1 2 Reset the alarm by pressing and releasing the POWER MODE button once after the CO reading falls below the low alarm point 3 4 2 Responding to an Over Range Alarm An over range alarm could indicate CO gas is above the detection limit of 1000 ppm The CO concentration reading is replaced by blinking brackets ΠΠΠΠ WARNING An over rang...

Page 14: ...the LCD Alarm LEDs flash once per second Backlight turns on Alternating high low tone chirps about twice per second Vibrator pulses once per second Over Range An over range condition indicates an extreme CO concentration Gas reading is replaced with a flashing ΠΠΠΠ Gas name and units flash OVER appears at the bottom of the LCD Alarm LEDs flash twice per second Backlight turns on Alternating high l...

Page 15: ...ntinue into Measuring Mode if the instrument must be used temporarily Figure 11 3 5 Alarm Setpoints 3 5 1 There are six alarm points The alarm points and their factory settings are summarized below There are no adjustments of alarm setpoints F S Full Scale Is triggered when CO concentrations go over the scale maximum of 1000 ppm WARNING Is triggered when CO concentrations reach 10 ppm which is the...

Page 16: ...he monitor is in Measuring Mode can result in death from carbon monoxide poisoning 4 2 1 Entering User Mode Begin with the monitor OFF 4 2 1 1 Press and hold the AIR button then press and hold the POWER MODE button Release both buttons as soon as the monitor beeps 4 2 2 Tips for Using User Mode To scroll from one menu item to the next press AIR and release it as soon as the alarm beeps To reverse ...

Page 17: ...he Date and Time Refer to Section 2 3 THIRTEEN PASS W When set to ON password required to enter User Mode Factory setting is OFF Do not adjust FOURTEEN ROM SUM Shows firmware version and firmware checksum Refer to Section 4 2 7 to display firmware FIFTEEN START Enters Measuring Mode from User Mode Press and release POWER MODE Figure 14 4 3 DISPLAY MODE Display mode is used to view and change setti...

Page 18: ...actory settings that should not be adjusted If operator accidentally enters an unfamiliar menu DO NOT PUSH BUTTONS THAT COULD CHANGE THE MONITORS FUNCTION Refer to Section 4 4 to return to factory settings SEQUENTIAL LIST OF DISPLAY MODE MENU ITEMS Number of times POWER MODE beeps is pressed and released to reach menu item Display Mode Menu Items Description Sections In Which Adjustments Are Found...

Page 19: ...e last 15 minutes There are no changes or resets on the STEL menu Figure 18 4 3 7 Displaying the TWA Screen 4 3 7 1 The TWA Screen displays the time weighted average TWA reading The TWA reading is the average reading over the last 8 hours If 8 hours have not elapsed since the last time the TWA reading was cleared the average is still calculated over 8 hours The missing readings are assigned a valu...

Page 20: ...in Maintenance Mode returns all parameters to their factory settings 4 4 3 1 Enter Maintenance Mode per Section 4 4 2 4 4 3 2 Press and release AIR 16 times and scroll to M DEF 4 4 3 3 Press and release POWER MODE The screen will alternate between M DEF YES MODE and NO AIR 4 4 3 4 Press and release POWER MODE to perform a default operation 4 4 3 5 The monitor will ask if you are sure you want to p...

Page 21: ...ection 5 7 the CMS 4 is ready for monitoring 5 2 2 Fresh Air Adjustment with Impurity Free 0 ppm CO Test Gas The following is required to set the fresh air baseline with impurity free test gas Impurity free test gas Stock No 11132 Calibration connector with 0 5 LPM fixed flow slide valve tubing and calibration cup 5 2 2 1 Make sure the monitor is off 5 2 2 2 Prepare the calibration connector regul...

Page 22: ... Figure 21 5 3 5 Screw the calibration connector with slide valve onto the test gas cylinder NOTE Test gas can be 25 ppm CO used to calibrate the monitor or it can be impurity free air used to set fresh air baseline when using impurity free air per Section 5 2 2 5 4 Connect Calibration Cup to Monitor Figure 22 5 4 1 Prepare the calibration connector with slide valve for calibration per Section 5 3...

Page 23: ...ollowing are required to bump test the CMS 4 Test gas with 25 ppm concentration of CO Stock No 25573 Calibration connector with 0 5 LPM fixed flow slide valve tubing and calibration cup Stock No 29767 assembled as noted in Section 5 3 5 6 1 Make sure the fresh air baseline was set per Section 5 2 5 6 2 Prepare the calibration connector regulator valve and 25 ppm CO test gas per Section 5 3 Make su...

Page 24: ...d as noted in Section 5 3 5 7 1 Make sure the fresh air baseline was set per Section 5 2 5 7 2 Prepare the calibration connector regulator valve and 25 ppm CO test gas per Section 5 3 Make sure the slide valve is closed 5 7 3 With the monitor OFF press and hold AIR and POWER MODE Release both buttons when the monitor beeps BUMP is displayed on the LCD The monitor is now in User Mode 5 7 4 Press an...

Page 25: ...operation and accurate readings of the monitor Maintenance should be performed by personnel properly trained for this work Improper service could result in malfunction and loss of life NOTE There are no serviceable parts within the CMS 4 case Do not remove the four screws holding the upper and lower case together or try to separate the case 6 1 Replacing the Lithium Battery Figures 27 and 28 The b...

Page 26: ...own in Figure 29 Before removing the sensor note the orientation of the old sensor alignment points as shown in Figure 30 The new sensor will insert with the same orientation Figure 29 NOTICE Do not force the sensor into the socket when correctly oriented the sensor easily fits into the socket Forcing it could damage the sensor or the socket 6 2 6 Carefully insert the replacement sensor Make sure ...

Page 27: ...filter It should be seated and recessed in the sensor gasket and should not overlap the edge of the sensor gasket 6 3 9 Clip the sensor retainer over the filters and sensor gasket and tighten its screw 6 4 Replacing the Hydrophobic Filter Figure 32 NOTE Replace the hydrophobic filter in a clean nonhazardous environment Replace the charcoal filter when replacing the hydrophobic filter 6 4 1 Make su...

Page 28: ... be recalibrated 1 Recalibrate the monitor 2 If the difficulties continue replace the sensor and calibrate The sensor may need replacement 1 Recalibrate the monitor 2 If the difficulties continue replace the sensor and calibrate Calibration fails The calibration gas values may not match the cylinder gas concentration 1 Make sure test gas and gas value are 25 ppm Refer to Sections 5 6 and 5 7 2 Mak...

Page 29: ... CLEMCO Figure 33 8 2 Monitor Replacement Parts Figure 34 NOTE There are no serviceable parts within the CMS 4 case Do not remove the four screws holding the upper and lower case together or try to separate the case Item Description Stock No 1 Sensor CMS 4 29768 2 Retainer sensor 29769 3 Retainer battery 29770 4 Gasket sensor 29771 5 Gasket battery retainer 29772 6 Filter hydrophobic each 29868 7 ...

Reviews: