Clemco CMS-4 Operation Manual Download Page 5

PREFACE 

 

 

 

[

 

IV

 

]

 

 

DAILY

 

SET

UP

 

CHECK

 

LIST

 

 

 

 
 
 
 

 

Make sure all blast operators are properly trained and suitably attired with a blast suit, safety boots, leather gloves, respiratory and hearing protection. Every day before start up, 
check all equipment components, including piping, fittings, and hoses, and valves, for leaks, tightness, and wear. Repair or replace as needed. Use the following checklist.

 

 

1.  PROPERLY-MAINTAINED AIR COMPRESSOR 

sized to provide 

sufficient volume (cfm) at given pressure for nozzle and other tools. ADD 50% 
volume (cfm) reserve to allow for nozzle wear. Use large compressor outlet and 
air hose (at least 4 times the nozzle orifice diameter). For oil-lubricated 
compressors, the employer shall use a high- temperature or carbon monoxide 
alarm, or both, to monitor carbon monoxide levels. If only high-temperature 
alarms are used, the air supply shall be monitored at intervals sufficient to 
prevent carbon monoxide in the breathing air from exceeding 10 ppm. Follow 
the manufacturer’s checklist and maintenance instructions.

 

2.  BREATHING-AIR COMPRESSOR 

(or oil-less ambient air pump) 

capable of providing Grade D quality air, located in a dust free area. Read # 1 
above.

 

3.  CLEAN, PROPERLY-MAINTAINED NIOSH-APPROVED SUPPLIED-AIR 

RESPIRATOR 

worn by blast operators, and other workers exposed to blast dust.  

Make sure all respirator components are in place — all lenses, inner collar, and 
cape. Thoroughly inspect all components for wear. The NIOSH approval (approval 
number is listed in the owner’s manual) is for a complete assembly from point of 
attachment on the CPF (sorbent bed) filter to the complete respirator. Substitution of 
any part voids the NIOSH approval. 

4.  CARBON MONOXIDE MONITOR/ALARM 

installed at the CPF filter or 

inside the supplied-air respirator for monitoring for the presence of deadly CO gas 
and warning the operator(s) when the CO level reaches an unacceptable level. 
When an ambient air pump is used for breathing air, a CO monitor provides a 
measure of safety. Read # 1 above.

 

5.  BREATHING-AIR FILTER (OSHA-REQUIRED 

sorbent bed filter) for 

removal of moisture and particulate matter in the compressed air breathing-air 
supply. Monitor the condition of the cartridge and replace when odor is detected or 
at 3 month intervals, whichever comes sooner. The breathing air filter does NOT 
detect or remove carbon monoxide (CO). Always install a CO monitor/alarm.

 

6.  BLAST MACHINE 

(bearing U or UM stamp, National Board Number, and 

Maximum Working Pressure) sized to hold a 30-minute abrasive supply. Examine 
pop-up valve for alignment. Check piping, fittings, screens, valves for tightness, 
leaks, and wear. Always ground the machine to eliminate hazard of static shock. 
Install a blast machine screen to keep out foreign objects. Use a blast machine 
cover if left outdoors overnight. Never exceed the maximum working pressure of the 
vessel.

 

7.  AIR LINE FILTER 

(moisture separator) installed as close as possible to the 

blast machine inlet and sized to match the size of the inlet piping or larger air supply 

line. Clean filter and drain often. Damp abrasive causes operational problems.

 

 

8.  REMOTE CONTROLS 

are required by OSHA and must be in perfect 

operating condition. Test and check all components to ensure all parts are present 

and fully functional. Use genuine replacement parts. NEVER mix parts from different 

manufacturers. Never use welding hose for remote control hose.

 

 

9.  BLAST HOSE 

should have an inside diameter sized to suit the blast nozzle. 

The ID should be three to four times the size of the nozzle orifice diameter. Blast hose 

should be arranged in as straight a line as possible from the blast machine to the 

work area, avoiding sharp bends.

 

 

10.  COUPLINGS AND NOZZLE HOLDERS 

should fit snugly on the hose and 

be installed with manufacturer recommended screws. Coupling lugs must snap firmly 

into locking position. Gasket must always be used to form a positive seal, and cotter 

pins must be installed. Replace gasket when wear, softness or distortion is detected. 

Check nozzle holder for thread wear; replace at any sign of wear. Install safety cables 

at all connections.

 

 

11.  NOZZLE 

orifice size should be checked and nozzle replaced when worn 

1/16” from original size. (No. 5 nozzle has 5/16” orifice diameter; replace when it 

measures 3/8”). Threads should be inspected daily for wear and nozzle should be 

replaced when wear is detected. Always use a nozzle washer.

 

 

12.  ABRASIVE 

must be a material specifically manufactured for blasting. It 

should be properly sized for the job. Check material safety data sheet for free-

silica, cyanide, arsenic, lead and other toxins and avoid use when these toxic, 

harmful substances are present.

 

 

SURFACE TO BE BLASTED 

should be examined for hazardous substances. 

Take appropriate protective measures as required by OSHA to ensure the blast 

operator, other workers in the vicinity, and any bystanders are properly protected.

 

 

©Clemco Industries Corp., Stock No. 20954P, 0692  Rev. F, 06/12 

 

1. Air Compressor

7. Air Line 
Moisture 
Separator

5. CPF Air Filter 

6. ASME Code 
Blast Machine

8. Remote 
Controls

9. Blast Hose

10. Hose Couplings and Safety Cables 

11. Appropriately Sized Nozzle 

12. Abrasive 

4. External or 
Helmet Mounted 
Carbon Monoxide 
Monitor /Alarm 

2. Breathing Air Compressor for 
High Pressure Respirators

 

3. NIOSH Approved 
Supplied-Air Respirator 

or Ambient Air Pump for 
Low Pressure Respirators

 

Summary of Contents for CMS-4

Page 1: ...om DATE OF ISSUE 07 2019 REVISION Do not use this equipment before READING this MANUAL and UNDERSTANDING its contents These WARNINGS are included for the health and safety of the operator and those in the immediate vicinity Failure to read and understand these warnings can result in injury or death Electronic files include a preface containing the same important information as in the orange cover ...

Page 2: ...ast equipment is only one component of an abrasive blasting job Other products such as air compressors air filters and receivers abrasives scaffolding hydraulic work platforms or booms equipment for lighting painting ventilating dehumidifying parts handling or specialized respirators or other equipment even if offered by Clemco may have been manufactured or supplied by others The information Clemc...

Page 3: ... other non breathable gas Never modify or change respirator air line connections without first testing the content of the line for safe breathing air Failure to test the line may result in death to the respirator user Breathing air quality must be at least Grade D as defined by the Compressed Gas Association specification G 7 1 per OSHA Regulation 29 CFR 1910 134 When compressed air is the breathi...

Page 4: ...terval clean all filters screens and alarm systems If spring loaded abrasive valves are used always cage spring before disassembly WARRANTY The following is in lieu of all warranties express implied or statutory and in no event shall seller or its agents successors nominees or assignees or either be liable for special or consequential damage arising out of a breach of warranty This warranty does n...

Page 5: ...ens valves for tightness leaks and wear Always ground the machine to eliminate hazard of static shock Install a blast machine screen to keep out foreign objects Use a blast machine cover if left outdoors overnight Never exceed the maximum working pressure of the vessel 7 AIR LINE FILTER moisture separator installed as close as possible to the blast machine inlet and sized to match the size of the ...

Page 6: ...ol to avoid possible injury or death NOTICE Notice indicates information that is considered important but not hazard related if not avoided could result in property damage CAUTION Caution indicates a hazardous situation that if not avoided could result in minor or moderate injury WARNING Warning indicates a hazardous situation that if not avoided could result in death or serious injury DANGER Dang...

Page 7: ...ion of Operation WARNING The CMS 4 detects carbon monoxide which can be life threatening When using the CMS 4 follow the instructions maintenance calibration schedule and warnings in this manual to assure proper and safe operation of the monitor and to minimize the risk of personal injury Carbon monoxide poisoning can result in death or serious injury 1 5 1 The CMS 4 respirator mounted carbon mono...

Page 8: ...ochemical cell that consists of two precious metal electrodes in a dilute acid electrolyte A gas permeable membrane covers the sensor face and allows gas to diffuse into the electrolyte The gas reacts in the sensor and produces a current proportional to the concentration of the target gas The current is amplified by the CMS 4 s circuitry converted to a measurement of gas concentration and displaye...

Page 9: ...ress and hold AIR b Immediately press POWER MODE and then release both buttons NOTE Each screen displays for 20 seconds If you do not press a button within 20 seconds the CMS 4 automatically returns to Measuring Mode Button Function POWER MODE Turns the monitor ON and OFF Activates Display Mode Enters instruction into CMS 4 s microprocessor Resets alarm after an alarm condition Also Turns the LCD ...

Page 10: ...RATION 2 1 Installation This section explains the initial installation of the CMS 4 to the inside of the respirator 2 1 1 Prepare Mounting Clip 2 1 1 1 Adhere Velcro loop to CMS 4 mounting clip Trim the Velcro loop strip to 3 4 x 1 Remove the backing and adhere it to the bottom of the monitors mounting clip as shown in Figure 4 2 1 2 Installing Monitor in Apollo 20 Respirator Refer to Section 2 1 ...

Page 11: ... the monitor when doffing the respirator at the end of the shift and to do any service or calibration including turning the monitor on and off Do this to make sure the monitor is fully functional before placing it inside the helmet WARNING Remove the monitor at the end of the shift turn it off and store it in a clean environment At the beginning of each shift turn on the monitor and make sure it i...

Page 12: ...air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 4 2 Turn on the monitor by pressing and holding the POWER MODE button until the alarm beeps 2 4 3 The monitor goes though the warm up sequence as noted in Paragraph 2 2 2 2 4 4 After the warm up sequence the monitor goes into Measuring Mode normal operating mode it beeps and the LCD displa...

Page 13: ...d to removing the respirator as soon as it is safe to do so 3 4 1 2 Reset the alarm by pressing and releasing the POWER MODE button once after the CO reading falls below the low alarm point 3 4 2 Responding to an Over Range Alarm An over range alarm could indicate CO gas is above the detection limit of 1000 ppm The CO concentration reading is replaced by blinking brackets ΠΠΠΠ WARNING An over rang...

Page 14: ...the LCD Alarm LEDs flash once per second Backlight turns on Alternating high low tone chirps about twice per second Vibrator pulses once per second Over Range An over range condition indicates an extreme CO concentration Gas reading is replaced with a flashing ΠΠΠΠ Gas name and units flash OVER appears at the bottom of the LCD Alarm LEDs flash twice per second Backlight turns on Alternating high l...

Page 15: ...ntinue into Measuring Mode if the instrument must be used temporarily Figure 11 3 5 Alarm Setpoints 3 5 1 There are six alarm points The alarm points and their factory settings are summarized below There are no adjustments of alarm setpoints F S Full Scale Is triggered when CO concentrations go over the scale maximum of 1000 ppm WARNING Is triggered when CO concentrations reach 10 ppm which is the...

Page 16: ...he monitor is in Measuring Mode can result in death from carbon monoxide poisoning 4 2 1 Entering User Mode Begin with the monitor OFF 4 2 1 1 Press and hold the AIR button then press and hold the POWER MODE button Release both buttons as soon as the monitor beeps 4 2 2 Tips for Using User Mode To scroll from one menu item to the next press AIR and release it as soon as the alarm beeps To reverse ...

Page 17: ...he Date and Time Refer to Section 2 3 THIRTEEN PASS W When set to ON password required to enter User Mode Factory setting is OFF Do not adjust FOURTEEN ROM SUM Shows firmware version and firmware checksum Refer to Section 4 2 7 to display firmware FIFTEEN START Enters Measuring Mode from User Mode Press and release POWER MODE Figure 14 4 3 DISPLAY MODE Display mode is used to view and change setti...

Page 18: ...actory settings that should not be adjusted If operator accidentally enters an unfamiliar menu DO NOT PUSH BUTTONS THAT COULD CHANGE THE MONITORS FUNCTION Refer to Section 4 4 to return to factory settings SEQUENTIAL LIST OF DISPLAY MODE MENU ITEMS Number of times POWER MODE beeps is pressed and released to reach menu item Display Mode Menu Items Description Sections In Which Adjustments Are Found...

Page 19: ...e last 15 minutes There are no changes or resets on the STEL menu Figure 18 4 3 7 Displaying the TWA Screen 4 3 7 1 The TWA Screen displays the time weighted average TWA reading The TWA reading is the average reading over the last 8 hours If 8 hours have not elapsed since the last time the TWA reading was cleared the average is still calculated over 8 hours The missing readings are assigned a valu...

Page 20: ...in Maintenance Mode returns all parameters to their factory settings 4 4 3 1 Enter Maintenance Mode per Section 4 4 2 4 4 3 2 Press and release AIR 16 times and scroll to M DEF 4 4 3 3 Press and release POWER MODE The screen will alternate between M DEF YES MODE and NO AIR 4 4 3 4 Press and release POWER MODE to perform a default operation 4 4 3 5 The monitor will ask if you are sure you want to p...

Page 21: ...ection 5 7 the CMS 4 is ready for monitoring 5 2 2 Fresh Air Adjustment with Impurity Free 0 ppm CO Test Gas The following is required to set the fresh air baseline with impurity free test gas Impurity free test gas Stock No 11132 Calibration connector with 0 5 LPM fixed flow slide valve tubing and calibration cup 5 2 2 1 Make sure the monitor is off 5 2 2 2 Prepare the calibration connector regul...

Page 22: ... Figure 21 5 3 5 Screw the calibration connector with slide valve onto the test gas cylinder NOTE Test gas can be 25 ppm CO used to calibrate the monitor or it can be impurity free air used to set fresh air baseline when using impurity free air per Section 5 2 2 5 4 Connect Calibration Cup to Monitor Figure 22 5 4 1 Prepare the calibration connector with slide valve for calibration per Section 5 3...

Page 23: ...ollowing are required to bump test the CMS 4 Test gas with 25 ppm concentration of CO Stock No 25573 Calibration connector with 0 5 LPM fixed flow slide valve tubing and calibration cup Stock No 29767 assembled as noted in Section 5 3 5 6 1 Make sure the fresh air baseline was set per Section 5 2 5 6 2 Prepare the calibration connector regulator valve and 25 ppm CO test gas per Section 5 3 Make su...

Page 24: ...d as noted in Section 5 3 5 7 1 Make sure the fresh air baseline was set per Section 5 2 5 7 2 Prepare the calibration connector regulator valve and 25 ppm CO test gas per Section 5 3 Make sure the slide valve is closed 5 7 3 With the monitor OFF press and hold AIR and POWER MODE Release both buttons when the monitor beeps BUMP is displayed on the LCD The monitor is now in User Mode 5 7 4 Press an...

Page 25: ...operation and accurate readings of the monitor Maintenance should be performed by personnel properly trained for this work Improper service could result in malfunction and loss of life NOTE There are no serviceable parts within the CMS 4 case Do not remove the four screws holding the upper and lower case together or try to separate the case 6 1 Replacing the Lithium Battery Figures 27 and 28 The b...

Page 26: ...own in Figure 29 Before removing the sensor note the orientation of the old sensor alignment points as shown in Figure 30 The new sensor will insert with the same orientation Figure 29 NOTICE Do not force the sensor into the socket when correctly oriented the sensor easily fits into the socket Forcing it could damage the sensor or the socket 6 2 6 Carefully insert the replacement sensor Make sure ...

Page 27: ...filter It should be seated and recessed in the sensor gasket and should not overlap the edge of the sensor gasket 6 3 9 Clip the sensor retainer over the filters and sensor gasket and tighten its screw 6 4 Replacing the Hydrophobic Filter Figure 32 NOTE Replace the hydrophobic filter in a clean nonhazardous environment Replace the charcoal filter when replacing the hydrophobic filter 6 4 1 Make su...

Page 28: ... be recalibrated 1 Recalibrate the monitor 2 If the difficulties continue replace the sensor and calibrate The sensor may need replacement 1 Recalibrate the monitor 2 If the difficulties continue replace the sensor and calibrate Calibration fails The calibration gas values may not match the cylinder gas concentration 1 Make sure test gas and gas value are 25 ppm Refer to Sections 5 6 and 5 7 2 Mak...

Page 29: ... CLEMCO Figure 33 8 2 Monitor Replacement Parts Figure 34 NOTE There are no serviceable parts within the CMS 4 case Do not remove the four screws holding the upper and lower case together or try to separate the case Item Description Stock No 1 Sensor CMS 4 29768 2 Retainer sensor 29769 3 Retainer battery 29770 4 Gasket sensor 29771 5 Gasket battery retainer 29772 6 Filter hydrophobic each 29868 7 ...

Reviews: