CMS-4 IN-HELMET CARBON MONOXIDE MONITOR
Page 17
© 2019 CLEMCO INDUSTRIES CORP.
www.clemcoindustries.com
Manual No. 29872
5.3
Prepare Calibration-Connector Regulator
Valve for Calibration ‒ Figure 21
5.3.1
The following materials are required to prepare
the calibration connector and test gas.
1. Calibration Test-Gas Cylinder (25 ppm CO)
Or
Impurity-Free Air Cylinder (used when setting
fresh-air baseline, per Section 5.2.2)
2. Calibration Connector with 0.5 LPM Fixed-Flow
Slide Valve
3. Calibration Tubing
4. Calibration Cup
OP E N
O
L
E
S
C
Figure 21
5.3.2
Slide one end of the 3/16" clear tubing onto the
barbed end of the calibration connector with slide valve.
5.3.3
Slide the other end of the 3/16" tubing onto the
inlet barb of the calibration cup.
NOTE: The calibration
cup (shaded in Figure 21) has directional flow. The
flow-direction arrow shows the direction of air flow.
The tube connects to the barb on the backend of the
flow arrow, as shown. The outlet barb (on the front end
of the flow arrow) remains open.
5.3.4
Make sure the slide valve is CLOSED by sliding
it toward the test gas cylinder, as shown in Figure 21.
5.3.5
Screw the calibration connector with slide valve
onto the test gas cylinder. NOTE: Test gas can be 25
ppm CO used to calibrate the monitor, or it can be
impurity-free air used to set fresh-air baseline when
using impurity-free air, per Section 5.2.2.
5.4
Connect Calibration Cup to Monitor ‒ Figure 22
5.4.1
Prepare the calibration connector with slide
valve for calibration, per Section 5.3. Make sure the
connector’s slide valve is closed.
5.4.2
Before attaching the calibration cup onto the
monitor, make sure it is correctly positioned, as shown in
Figure 22. The cup’s flow arrow must point up and to the
left, as shown.
CLEMCO
Figure 22
5.4.3
To attach the cup, refer to Figure 23 and hook
the calibration cup’s lower retaining clip under the lower
slot, as shown in Figure 23.
Flow
Arro
w
Figure 23
5.4.4
Gently press the other end of the cup onto the
monitor until it snaps into place, as shown in Figure 24.
Calibration Cup
Lower Clip
Clip Slot
Tubing
View from sensor-end of monitor
Clip Slot
Upper Clip
Sensor Retainer
CMS-4 Monitor
Open End of Calibration Cup
3/16" ID Tubing
Upper Retaining Clip
Lower Retaining Clip
Test Gas Cylinder
Calibration Connector with
Fixed-Flow Slide Valve
Slide Valve
Calibration Cup
3/16" ID Tubing
Flow arrow points away
from calibration tubing.
OPEN
CLOSE