CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
51
•
Take the HEPA filter out of the cabinet and put it into an appropriate plastic bag;
•
Seal the plastic bag airtight;
•
Remove the gloves and put them in the appropriate container.
Position the new HEPA filter
•
Unpack the new filter, check specifications and register the serial number and specifications;
•
Filter certificate has to be stored on a safe place;
•
Put Vaseline on the seal of the HEPA filter (both sides);
•
Position the HEPA filter (take care for possible damage of the sealings!);
•
Place the new HEPA-filter;
•
Put the cabinet together in the reverse way as described above;
•
Put of your gloves and put these in an appropriate bag;
•
Do an efficiency test according to the procedure described in Appendix III.
The valid procedures for the disposal of possible contaminated material must always be followed.
When the service engineer thinks this is not enough this will be discussed with the responsible
person so that enough precautions can be made.