CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
39
Product protection becomes less, there is a chance that aerosols from the outside go into the
working area.
Flow too high
Airflow too high
Figure 7
Product protecting crossflow is too high.
Product protection stays intact.
6.2
Other alarms
Other alarms are:
Fan failure
Fan failure
Figure 8
Most likely the fuse is broken. Replace it with a new one. If the alarm is still there, contact your
service organization.
To reset this alarm you must disconnect the power plug from the main voltage.
Communication error
Communication error
Figure 9
Contact your service organization.
Overview possible alarms
No alarm
Flow too low
Flow too high
Product protection:
Yes
No
Yes
Table 1: Alarm situations in work position
Procedure in case of alarms
In case the alarm is still present:
Stop working, and push
button
or the standard
operation procedure has to be
followed.
The cabinet goes into the saving position.