CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
11
Appendix I: Support frame CLF on page 49 for an explaining drawing):
•
Mount the support yokes (Pos. 5) to the reinforcement plate (Pos. 6);
•
Mount the feet support (Pos. 7);
•
Position the covers (Pos. 4) and adjusting feet (Pos. 3).
Placing the cabinet
•
Place the cabinet right on the forks of a forklift truck or lifting platform;
DANGER BY INDUSTRIAL VEHICLE
Be careful: Make sure the cabinet cannot slide from
the forks.
•
Move the cabinet to the correct height and position, perform this action as precise as possible
above the support frame;
•
The weight must be equally divided between the both forks and the cabinet must be horizontal
levelled;
•
The cabinet should be aligned with the support frame.
•
Mount the cabinet to the support frame by means of the screws.
•
Connect the powerplug to a grounded socket that is easy accessible.
In case of maintenance the cabinet must be completely free of voltage. This means that the
connector in the socket must be easy to reach, otherwise a main-switch must be mounted
between the cabinet and the power supply so that the cabinet can be cut off of the power-circuit.
3.2
Transport instructions
Before moving the cabinet make sure the power cable is disconnected of the power supply
The cabinet must be transported on a strong solid pallet.
After every transport the cabinet must be tested and calibrated.
GENERAL DANGER
Be careful: High center of gravity, crossover point.
3.3
Dismantling & Disposal
For dismantling and disposal follow the next instructions:
•
See
§7.1 Cleaning the installation on page 41;
•
See
§8.2 Replacement of the pre-filter on page 42
;
•
See
Appendix II: Replacement of the HEPA-filters
on page 50;
•
Break the unit down into its component parts. You can re-cycle these components parts,
dispose of them in accordance with local requirements.
For advice or information about dismantling or disposal please contact Clean Air Techniek B.V.