CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
44
8.4
Service contact
For advice or any technical questions please contact your service organisation. Please refer to the
following items:
•
Type, Class, Project number and Serial number mentioned on the type plate
(for the position of the type plate see
page 12
and
page 57
);
•
Version number of the software used in the cabinet.
Software Version
1
Cabinet in work mode.
See
§5.3 Starting up the cabinet
on page 19.
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off
2
Software version:
•
Push button
(Menu).
•
Push button
(C).
•
Push 2x button
(=>).
Factory Adjustments
Version: X.X xxxxx I/O: VXXXX OPT: VX.X
>>
+
-
Esc
8.5
Service contract
For information and orders concerning service or service contracts please contact Clean Air
Techniek B.V.
8.6
Testing
Recommended tests for first installation or after changing the HEPA filters are:
-
Velocity tests (see
Appendix IV Measuring method air velocities
on page 53);
-
Filter Integrity tests (see Appendix III: Efficiency test of the HEPA-filters
on page
52
);
-
Alarm tests (see Appendix VI: Alarm-test procedure
on page
55
).
Recommended tests after installing with a regular interval are:
-
Velocity tests (see
Appendix IV Measuring method air velocities
on page 53);
-
Filter Integrity tests (see Appendix III: Efficiency test of the HEPA-filters
on page
52
);
-
Alarm tests (see Appendix VI: Alarm-test procedure
on page
55
).