[4] Function 

<Normal mode> 

Item Description 

Flow rate display  Instantaneous flow rate is displayed. 

Integrating function  Integrated flow rate is displayed. 

Peak hold 

function 

Maximum and minimum flow rate values during the specified 
period can be read. 

Set-point 

verification 

Switch data (set-point value and operation mode) are 
displayed to check. 

Key lock 

Setting changes are disabled to prevent incorrect operations. 

Error display 

function 

The error state is displayed. 

 

<Standard setting mode> 

Item Description 

Switch output 

Having 2 pieces of switch output, 7 operation patterns and stop 
of operation can be set. 

Forcible output 

function 

Switch output is turned on forcibly to check wiring connection 
and initial operation of input unit.   

0 point adjustment  Deviation of the display from o is corrected. 

 

<Detailed setting mode> 

Item Description 

Flow direction 

selection 

Only bi-directional type, flow direction can be switched. 

CH2 function 

selection 

Sets the CH2 function. 
Selects “Switch output”, ”External input of auto reference”, 
or ”External input of integrated reset”. 

Auto reference 

function 

When CH2 function selected external input of auto reference, 
setting value of switch output can be taken by external input or 
key operation. 

Response time 

setting 

Sets the response time. 
The response time can be selected from 20ms to 1280ms.   

Display speed 

selection 

Change the speed of the displayed. 

Sub-display 

selection 

Change the indication of the sub-display. 
Selects “Flow direction”, ”Flow rate unit”, or “Working fluid”. 

Displayed color 

selection 

Displayed color can be changed. 

Hysteresis fixed 

value selection 

Sets hysteresis of the window operation mode and the auto 
reference mode. (8 steps) 

Flow rate unit 

selection 

Flow rate unit can be changed. 
Standard condition (ANR): Converted to volumetric flow at 20°C

and 1 atmospheric pressure (101kPa). 

Reference condition (NOR): Converted to volumetric flow at 0°C

and 1 atmospheric pressure (101kPa).(calculation value) 

Eco mode setting  Current consumption can be lowered. 

When the product is left for 1 min. without any operation, it’s 
shift to eco mode. 

Reset setting 

Return to default settings (factory settings) 

 

[5]How to operate 
5-1.Normal mode

 

<

 

Displaying the integrated flow > 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Note: Integration is reset with the external input. See auto reference function. 
Note: Integration is also reset or Integrating function display setting is keeping when power is 
turned OFF.

 

 

 

<Peak hold function>

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

<Set-point verification method>

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

<

 

Key lock function>

 

Key lock 

 
 
 
 
 
 
 
 
 
 

 

key press After key lock and It is displayed “Loc”. 

Key unlock

 

 
 
 
 
 
 
 
 
 
 
 

Note: Keys are unlocked when the controller is shipped. Lock keys if necessary. 

The key lock/unlock state is held even if power is turned OFF. 
While key lock, all the operations are not accepted excluding the key lock 
release operation. 
While key lock, If the key is operated, it becomes a “Loc” display. 

mL

 

/

 

min

 

 

 

mL

 

/

 

min

 

 

 

mL

 

 

Display unit selection 

Instantaneous flow rate display   
(unit:: mL/min (or) L/min) 

Integrated flow display (unit: mL (or) L)

,<Instantaneous flow rate display> 

Press once. 

Press once 
(determination). 

mL

 

 

<Integrated flow display>

 

 

Integration reset 

Integration is reset when the        and 
      keys are held down for 2 sec.

 

 

Press simultane-
ously for 2 sec.

mL

 

/

 

min

 

 

 

CH1 data display 

<Instantaneous flow rate display> 

Being held down

 

To instantaneous flow rate display 

Separate 

CH2 data display 

Lower limit

 

Upper limit

 

Operation waveform 

Displays in turn 

mL

 

/

 

min

 

 

 

mL

 

/

 

min

 

 

 

When CH1 selected window operation[1] 

When CH2 selected external input of 
integrated reset. 

Separate 

Being held down

 

<Instantaneous flow rate display> 

mL

 

/

 

min

 

 

 

Peak hold display 

and

     

 press simultaneously 

mL

 

/

 

min

 

 

 

mL

 

/

 

min

 

 

 

mL

 

/

 

min

 

 

 

Being held down 

Being held down 

<Peak value displayed>

<Bottom value displayed> 

Press once. 

Reset peak hold function.

 

To instantaneous flow rate display 

mL

 

/

 

min

 

 

 

<Instantaneous flow rate display (key unlock)> 

and

 

   

 

  press simultaneously for 1 sec. 

mL

 

/

 

min

 

 

<Instantaneous flow rate display (key lock)> 

mL

 

/

 

min

 

 

 

<Instantaneous flow rate display (key lock)> 

mL

 

/

 

min

 

 

 

<Instantaneous flow rate display (key unlock)>

and

 

   

 

  press simultaneously for 3 sec. 

-3- 

Discontinue

Summary of Contents for Rapiflow FSM2-N Series

Page 1: ...M2 N series Integrated needle valve type Read this instruction manual carefully before using this product particularly the section describing safety Retain this instruction manual with the product for further consultation whenever necessary 2 n d E D I T I O N Discontinue ...

Page 2: ...vided but damage could result if the needle is turned too far Power supply voltage and outputs must be used with the specified voltage Applying the voltage more than specified voltage may cause malfunction damage of sensor electric shock or fire Do not apply load more than the rated output Damage or fire of the output may be caused For wiring stop control unit machinery and equipment and turn off ...

Page 3: ...2 function selected switch output CH2 function selected external input PNP output Model no FSM2 P CH2 function selected switch output CH2 function selected external input 1 2 3 The panel bessel is inserted from the front side of the panel The panel Bessel is inserted from the back side of the panel and holding with the screw The separation indicator is inserted from the back side of the panel 4 5 ...

Page 4: ...ure 101kPa calculation value Eco mode setting Current consumption can be lowered When the product is left for 1 min without any operation it s shift to eco mode Reset setting Return to default settings factory settings 5 How to operate 5 1 Normal mode Displaying the integrated flow Note Integration is reset with the external input See auto reference function Note Integration is also reset or Integ...

Page 5: ...steresis operation 1 Hysteresis operation 2 Integrated output 1 Integrated output 2 Integrated pulse CH1 operation mode change Press once determination CH1 ON OFF data setting mL min Lower limit data setting Upper limit data setting Count UP Count down Press once determination CH2 operation mode change Set same as CH1 CH2 ON OFF data setting Set same as CH1 Press once determination Press once dete...

Page 6: ...lue At instantaneous flow rate display press key for 8 sec to enter to detailed setting mode To auto reference setting display To response time setting display To display speed setting display To sub display setting display To displayed color setting display To hysteresis fixed value setting display To flow rate unit setting display To eco mode setting display To reset setting display To model num...

Page 7: ...display To instantaneous flow rate display Press once determination Input point mL min Lower limit mL min Upper limit mL min Initial value 0 0 1st 123 0 123 2nd 234 123 234 3rd 45 45 234 4th 345 45 345 5th 456 345 456 Display speed setting display Setting of display speed 250msec 500msec 1000msec To instantaneous flow rate display Press once determination Press or key to select display speed Press...

Page 8: ... by noise Keep FSM2 main body and cable away from the noise source Sensor chip is broken Replace FSM2 Foreign matter sticking to sensor chip Replace FSM2 Poor precision Malfunction caused by noise Keep FSM2 main body and cable away from the noise source Reduce pulsation by installing tank etc Change the response time Change the display speed Pulsation of air Fault in power source not enough voltag...

Page 9: ...splay Dual 2 4 digit 7 segm ent Two color LCD Flow rate range 3 F 0 to 500 mL min 0 to 1000 mL min 0 to 2 00 L m in 0 to 5 00 L m in 0 to 10 00 L m in 0 to 20 0 L m in 0 to 50 0 L m in 0 to 100 0 L m in 0 to 200 L m in Flow rate display Note 1 2 Display resolution 1m L m in 0 01L m in 0 1L m in 1L m in Flow rate range 9999999m L 99999 99L 999999 9L 9999999L Display resolution 1m L 0 01L 0 1L 1L In...

Page 10: ...e8 50m A or less Lead wire φ3 7 AW G26 5 Functions Flow rate display flow rate display peak hold switch output and analog output Outside diam eter of insulator is Φ1 0 Installation attitude Both vertical and horizontal Strait piping section Not required Protective structure Note 9 IEC standards IP40 Protective circuit Note5 Power supply and switch output reverse connection protections and switch o...

Page 11: ...o FSM2 LB1 Model No FSM2 KHS N How to panel mount Panel hold matching Fig Plate thickness t0 8 to 6 Single mount Sticking mount 2 M3depth5 Model No Port size Dimension A FSM2 H04 Φ4 push in 89 FSM2 H06 Φ6 push in 90 Model No Port size Dimension A FSM2 H08 Φ8 push in 101 6 FSM2 H10 Φ10 push in 113 1 2 M3depth5 A B Model No Dimension A Dimension B FSM2 H04 H06 S06 N 28 5 40 5 FSM2 H08 H10 S08 N 30 0...

Page 12: ...ncluded H10 Φ4 push in resin K Inspection results included S06 Rc1 8 stainless S08 Rc1 4 stainless 10 Needle valve 11 Clean N Needle valve Blank none P70 Particle occurrence prevention P80 Oil treatment prohibited Combination of flow rate range port size and working fluid 5 Port size body material H04 H06 H08 H10 S06 S08 005 010 020 050 100 200 500 101 4 Flow rate range 201 Table for clean specifi...